− 21 −
Maintenance Inspection
■
Daily Inspection
■
Regular Inspection
In order to maintain the accuracy of the Gas Detector, request a regular inspection from your New-
Cosmos Electric representative.
All XP-3000 series gas detectors which are calibrated have the same algorithm to calculate the
measurement data to the concentration.
■
Main Replacement Parts
6. Maintenance Inspection
Check item
Description
Gas sampling hose
Make sure that the gas sampling hose has no bends or damage
and that it is securely connected.
Drain filter
If the filter element in the drain filter is dirty or wet or water
remains in the drain filter, clean the inside of the drain filter (and
1-m-long gas sampling hose) and replace the filter element with
a new one (see page 19).
Remaining battery level
If the remaining battery level is low, replace the batteries with
new ones (see page 20).
Airtightness
Press the power switch ON to display the gas concentration
screen, cover the gas sampling pipe tip with a finger and check
the error message "Err-P"(blockage malfunction) is displayed.
As "Err-P" is displayed, it is a normal state and operable by
pressing the power switch ON again. if "Err-P" is not displayed
the damage of the gas sampling pipe or the airtight defect of the
sensor packing or pump diaphragm may be considered.
Contact your New-Cosmos Electric representative for repairs or
replacements.
Product
Model
Remarks
Filter Element
FE-2 (10 filters)
Содержание XP-3110
Страница 30: ...MEMO...
Страница 31: ......