Operating Procedure

3. Operating Procedure

 Procedure

1. Inserting Batteries

The Gas Detector is delivered without batteries inserted. Refer to pages 20 to insert the provided 
batteries.

2.  Turning Gas Detector ON 

 Warming Up 

 Displaying Gas Concentration 

Screen

(1)  Press the POWER switch. The buzzer will beep and the Gas 

Detector will be turned ON.

(2)  The LCD main screen will display Adj and the bar graph will 

count down while the Gas Detector is warming up. The LCD 
sub screen will display Time.

(3)  When the sensor is stabilized, the buzzer will beep and the 

gas concentration screen will be displayed.

WARNING

Be sure to place the connected gas sampling hose in clean air when 
turning ON the Gas Detector. Zero calibration is done automatically. 
The wrong gas concentration will be displayed if zero adjustment is 
done in gas atmosphere.

WARNING

Be  sure  to  conduct  a  daily  inspection  before 
performing detection work (see page 21).

CAUTION

Even after displaying the gas concentration screen, zero indication 
might be unsettled and flash. In that case, wait for three minutes and 
conduct zero adjustment (refer to page 9).

Note

If an error message is displayed, refer to Error Messages on page 18.

Procedure

1

Insert

the

batteries.

2

Turn  ON 

t he   Ga s 

Detector.

Displays the gas 

concentration 

screen.

Warms 

up.

3

Detects

4

Turn OFF the 

Gas Detector

Содержание XP-3110

Страница 1: ...listed on the left Please keep this instruction manual available for quick reference when needed Before use carefully read this instruction manual and fully understand the content XP 3110 Combustible...

Страница 2: ...arm Buzzer Silence Settings 13 Changing the Sample Gas 14 Logging 15 Data Logger Communications Screen 16 Time Settings 17 4 Error Message Malfunction Alarm 18 5 Replacement of Consumable Parts 19 Fil...

Страница 3: ...eration has not been confirmed in other OS versions Hard disk Free space of 6 Mbytes or more CD ROM drive A CD ROM drive capable of reading CD R data The software is provided in a CD R disc USB port A...

Страница 4: ...g the Gas Detector to wear antistatic clothing and electrically conductive shoes antistatic work shoes The floor should also be electrically conductive with an electrical leakage resistance of 10 M or...

Страница 5: ...has an explosion proof construction Do not disassemble or modify the Gas Detector or change the structure or electric circuits of the Gas Detector Otherwise the explosion proof performance of the Gas...

Страница 6: ...ampling pipe connection port Connects to the gas sampling hose 3 Exhaust port Exhausts sampled gas 4 Buzzer opening Sounds a buzzer 5 Gas type seal Displays the model number and the types of detectabl...

Страница 7: ...range of the bar graph 10 LCD sub screen Displays gas concentration and messages J Indicates time K Displays a gas number confi rmed if the number of detectable combustible gases is two or more 11 Co...

Страница 8: ...k pocket Used to hold the sampling hose or provided filter No Name Function 1 Gas sampling hose Leads gas to the Gas Detector 1 m long 2 Coupler Connects to the Gas Detector 3 Drain filter DF 4 Preven...

Страница 9: ...ntration screen will be displayed WARNING Be sure to place the connected gas sampling hose in clean air when turning ON the Gas Detector Zero calibration is done automatically The wrong gas concentrat...

Страница 10: ...ressing the BZ STOP switch The screen will return to the clock display in approximately three seconds To change the sample gas refer to page 14 The gas number however cannot be confirmed during a gas...

Страница 11: ...sure to conduct zero adjustment or the precise measurement of gas may not be possible Range To change the bar graph range press the POWER switch while the gas concentration screen is displayed The ra...

Страница 12: ...oncentration reaches the alarm level again the gas alarm buzzer will sound again If the pointer scales out when detecting high concentration gas OL will be displayed High concentration gas may have an...

Страница 13: ...able gas selection Allows detectable gas selection if the number of detectable combustible gases is two or more P 14 4 LCD selection This detector can not be set 5 Execution and completion of logging...

Страница 14: ...lumination to flashing 3 Press the LIGHT switch to change from OFF to ON 4 Press the MENU switch to enter the settings The ON indication will change from flashing to steady illumination and 1PE will c...

Страница 15: ...dy illumination to flashing 3 Press the BZ STOP switch and change ON to OFF will change to 4 Press the MENU switch to enter the settings The OFF indication will then change from flashing to steady ill...

Страница 16: ...ing and 1 01 will start flashing 3 Press the LIGHT switch to select the gas number 4 Press the MENU switch to enter the settings The 1 02 indication will then change from flashing to steady illuminati...

Страница 17: ...d OFF will change from steady illumination to flashing 3 Press the LIGHT switch to change OFF to ON 4 Press the MENU switch to enter the settings The ON indication will then change from flashing to st...

Страница 18: ...op flashing and ON will start flashing Data logger communication with the PC will then be possible Operate the PC for data logger communication 4 When communication is finished press the POWER switch...

Страница 19: ...the setting and press the MENU switch to enter the value 5 Date Setting When the MENU switch is pressed CL3 will change from flashing to steady illumination and the date value will change from steady...

Страница 20: ...the battery voltage drops and the remaining battery level is low The Gas Detector will not work when the battery voltage is low Replace the batteries see pages 20 Use approved batteries Duracell MN150...

Страница 21: ...2 Use a small flat blade screwdriver to remove the O ring in the filter case 3 Replace the FE 2 Filter Element with a new one and reassemble the parts WARNING If water is drawn into the inside of the...

Страница 22: ...into the case part No 3620584053 and zip it shut before entering the hazardous location WARNING Use Panasonic LR6XJK Toshiba LR6 or Duracell MN1500 batteries Use of any other manufacturer or type wil...

Страница 23: ...in filter clean the inside of the drain filter and 1 m long gas sampling hose and replace the filter element with a new one see page 19 Remaining battery level If the remaining battery level is low re...

Страница 24: ...rned on even if you press the POWER switch T h e p o l a r i t y i s incorrect Install batteries again in the correct polarity Battery r e p l a c e m e n t P20 Battery life Replace the batteries The...

Страница 25: ...mos s sole option replacement or repair of such non conforming product or refund of the original purchase price of the non conforming product In no event will New Cosmos be liable for any other specia...

Страница 26: ...lkaline batteries Approx 30 hrs Alkaline batteries Approx 20 hrs Approx 15 hrs Methane specifications At an ambient temperature of 20 C with the alarm backlight and data logging function turned OFF Op...

Страница 27: ...Rev 2009 12 08 UL 60079 11 5th Ed Rev 2011 05 05 UL 60079 1 6th Ed Rev 2009 04 10 EN 60079 0 2012 EN 60079 0 2012 EN 60079 11 2012 EN 60079 1 2007 IEC 60079 0 6th Ed Corr 1 IEC 60079 0 6th Ed Corr 1...

Страница 28: ...d the resistance of the platinum filament will decrease This resistance change is output as a voltage deviation change of the bridge circuit The sensor is very sensitive to low concentrations making i...

Страница 29: ...sphere are to be placed in an enclosure which can withstand the pressure developed during internal explosion of an explosive mixture and which prevents the transmission of the explosion to the explosi...

Страница 30: ...MEMO...

Страница 31: ......

Страница 32: ...his instruction manual is lost contact New Cosmos Electric or the nearest branch or sales office shown below A copy of the instruction manual will be delivered to you A fee is charged XP 3000ATEXETC 0...

Отзывы: