2-47
1010PVNFM-3A
Section 2
Max Viscosity cS
The [Max Viscosity cS] menu cell stores the maximum range for the analog span. The viscosity
units are centistokes with no alternatives offered.
Min Viscosity cS
The [Min Viscosity cS] menu cell stores the minimum range for the analog span. The viscosity
units are centistokes with no alternatives offered.
Max Temperature
The [Max Temperature] menu cell stores the maximum temperature range for the analog
span. If metric preferences are selected
prior to site creation
the temperature units will be in
Centigrade. If English preferences are selected
prior to site creation
the temperature units
will be in Fahrenheit.
Min Temperature
The [Min Temperature] menu cell stores the minimum temperature range for the analog span.
If metric preferences are selected
prior to site creation
the temperature units will be in Centi-
grade. If English preferences are selected
prior to site creation
the temperature units will be
in Fahrenheit.
2.7.2 SET ALARM LEVELS
The Set Alarm Levels menu allows you to select system alarm functions. Alarms appear lo-
cally on the LCD digital display. In addition, you can use the Relay Setup menu to assign those
functions to the system’s relays. You may select from high or low energy or flow rate, liquid
interface (Vs) and liquid aeration alarm functions. Entry of all alarm setpoints is accomplished
using the selected rate units. You can enable or disable a Makeup Alarm Latch to keep the
makeup alarm active until you reset it manually by an <Fn>6 simultaneous key press.
To access the Alarm Setpoints menu press
High Flow
The [High Flow] menu cell stores the maximum range for the analog span.
To set the High Flow alarm value:
To activate numeric entry press .
Use the numeric keys to type the High Flow value.
To store the data press .
Low Flow
The [Low Flow] menu cell stores the maximum range for the analog span.
To set the Low Flow alarm value:
To activate numeric entry press .
Use the numeric keys to type the Low Flow value.
To store the data press .
ENT
ENT