2.11.7 TROUBLESHOOTING TIPS
The System 1010 has highly reliable circuitry and will provide trouble-free operation within
specified environments. Even a well-crafted precision instrument can fail if exposed to ex-
treme temperature or vibration conditions during service, storage or transportation. The
Diagnostic menu shows how the system interprets a problem. The test functions and alarm
indicators identify “hidden” problems automatically. If a problem seems unsolvable, call our
Technical Service Department or your local Controlotron Representative for expert help.
Flow Computer Messages
Certain actions or conditions invoke messages that may appear as a pop-up window, in the
right-hand column of a given menu cell, or the highlighted prompt line at the top of the dis-
play screen. The following is a list of messages that you may encounter along with explana-
tions, and in some cases, a recommended action.
MESSAGE
DESCRIPTION
Memory Full!
Response to an attempt to save site data, when data memory is full.
Delete an obsolete site or clear Datalogger memory to make room
for the new data.
Memory Corrupted! Memory read error occurred while accessing the active site data.
Chan Not Setup
Response to an attempt to invoke an operation that requires a
channel to be enabled. Enable the channel
[Channel
Setup
Channel Enable
Yes
]. Note that a channel cannot be enabled un-
til its transducers are operating.
Clr All Data?
Response to pressing the F4 key. Use the F4 function to restore
operation if a severe event (e.g., a violent power surge) disrupts
system operation.
Clr Dynamic Memory? Response to pressing the F4 key, then selecting [No] for Clr Saved
Data?
<EOT>
Response to a request to output datalogger data to the printer or
the Graphics screen when no datalogger data exists. Set up the
Datalogger.
Empty Pipe - Press
Prompt to empty the pipe during the Actual MTY procedure. After
[ENT]
emptying pipe, press <ENT>.
Fill Pipe - Press [ENT] Prompt to fill the pipe during the Actual MTY procedure. After fill-
ing the pipe, press [ENT] (see paragraph 3.1.3).
No Sites
Response while trying to recall/delete a site setup when no Sites
are stored.
Not Installed
Response to an attempt to access a menu function that is not in-
cluded.
Re-space Index
Upon measuring the liquid sonic velocity (Vs), the meter recom-
mends re-spacing the transducers to improve performance.
Invalid Setup
During the Initial Makeup, the system detects invalid transducer
spacing, erroneous liquid/pipe parameters, or some other factor
that prevents it from completing the Initial Makeup. This may be
due to one of the following:
• An out-of-range data entry.
• An invalid condition; (e.g., overlapping transducers in Reflect
Mode). If selecting Direct Mode does not resolve, review all
(use Direct Mount)
è
è
(continued)