2-73
1010PVNFM-3A
Section 2
“Manual” Adjustment Procedure
l
In the [Site Setup Data] Menu, press the <Down Arrow> and scroll to the [HF] menu
cell. Press the <Right Arrow> and a pop-up [Manual] prompt will appear as shown
below.
Note: Press the <Up/Down Arrow> to select [Automatic], if desired.
l
Use the numerical keys to input the desired correction value. Press <ENT> to input
value.
l
In the [Diagnostic Data] Menu, highlight [Path Select] and select the desired transducer
path. Press <ENT> to select path.
l
Press the <Down Arrow> and scroll to the [Site Setup Data] menu cell. Press the <Right
Arrow> to select it.
l
The new correction value will appear next to the [HF] menu cell as shown below.
“Automatic” Adjustment Procedure
l
In the [Site Setup Data] Menu, press the <Down Arrow> and scroll to the [HF] menu
cell. Press the <Right Arrow> and a pop-up [Manual] prompt will appear.
l
Press the <Up or Down Arrow> to select [Automatic] then press <ENT>.
fx (drive)
29
N (burst length)
5
Ltn in
5.316
Vf max CU
>Manual
Vs max m/s
= -0.120
Vs min m/s
1355.00
Empty %
Samples/Cycle
29
HF
>-0.000
Controlotron
Dual Path
4SS10G
Site Setup Data
fx (drive)
29
N (burst length)
5
Ltn in
5.316
Vf max CU
Adjustment
Vs max m/s
>Manual
Vs min m/s
1355.00
Empty %
Samples/Cycle
29
HF
>-0.000
Controlotron
Dual Path
4SS10G
Site Setup Data
fx (drive)
29
N (burst length)
5
Ltn in
5.316
Vf max CU
Vs max m/s
Vs min m/s
1355.00
Empty %
Samples/Cycle
29
HF
>-0.120
Controlotron
Dual Path
4SS10G
Site Setup Data