![Controlotron 1010PVDN Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/controlotron/1010pvdn/1010pvdn_field-manual_2658905046.webp)
2-14
1010PVNFM-3A
Section 2
To select a Liquid Class:
To access the option list press
Scroll through the list to find the closest match to the application’s liquid.
To register selection press
How to Edit the Estimated Vs (liquid sonic velocity)
During transducer installation, the flow computer bases its initial transducer spacing recom-
mendation on the value stored in this menu cell. Estimated Vs m/s allows you to review and
modify (if necessary) the Vs value for the liquid class you selected.
After you install the transducers, the computer will measure Vs directly. If the displayed
(or edited) Vs stored in this menu cell is accurate, the transducers will be spaced correctly.
This will eliminate the need to re-space the transducers after their initial installation. How-
ever, if the estimated Vs is substantially different from the measured value, the computer will
request you to re-space the transducers during the transducer install procedure.
NOTE: During the Xdcr install procedure (see paragraph 2.4), you can ignore a
respacing request by pressing the <Down Arrow> instead of <ENT>. If this
triggers a Spacing Alarm, check [Diagnostic Data/Site Setup Data] to make
sure the measured Vs value does not exceed the Vs max or Vs min items.
To edit the Estimated Vs m/s:
To activate numeric entry press
Type the sonic velocity value using meters-per-second.
To register the data press
ENT
ENT
2.3.1 HOW TO SELECT A LIQUID CLASS
The liquids listed in the Liquid Class option list are representative samples of the class of
liquids to which they belong. Selecting a named liquid fills in the Estimated Vs (m/s), Viscos-
ity (cS), and Density (SG) parameters automatically. However, you have the option of editing
these parameters individually to fine tune the liquid settings. If you do not find a liquid that
matches your application, then you can select [Other]. This selection will not provide a liquid
name or automatic parameter entry.
Liquid Class
Controlotron Dual Path
SITE 1
Access Liquid Option List
Estimated Vs m/s
1482
Viscosity cS
1.00
Density S.G.
1.000
Select Liquid
Water 20C/68F
Use this menu cell to edit the
Estimated Vs ONLY if you are
sure of a more accurate value.
This will reduce the chances of
a re-spacing request for clamp-
on flow transducers during the
install procedure.
Use this menu cell to edit
the liquid viscosity ONLY
if you are certain that the
data you enter is
accurate. An erroneous
entry could affect meter
calibration.
automatically enters
Use this menu cell to
edit the specific
gravity of the liquid.
Содержание 1010PVDN
Страница 185: ...APPENDIX A Couplant Installation Instructions...