3-6
1010PVNFM-3A
Section 3
Chain Tension
Transducers
Chain
PIPE
VIEW FROM
TOP OF PIPE
INSTALLATION - DIRECT MOUNT WITH EZ CLAMP AND SPACER BAR
1. Perform all required menu steps until the flow computer issues the number index and
prompts you to press [ENT] to finish the transducer install routine. Stop at this point. Note
the number index value displayed in the Pick/Install menu. You will use this number to
properly space the transducers. Check to ensure that you have a matched set of transduc-
ers. They both should have the same S/N number but marked with either an “A” or “B” (e.g.,
100A and 100B).
2. Temporarily position one transducer (in the 9 o’clock position) at the location where you
have determined it would be mounted. Ensure that this is a smooth area without any raised
spots (seams, etc.). Mark a generous area around the transducer where it contacts the pipe
(with a pencil or chalk). Remove the transducer.
3. Prepare the area you marked by de-greasing the surface if needed and removing any grit,
corrosion, rust, loose paint or surface irregularities with the abrasive pipe conditioning
material provided.
4. Attach the EZ clamp to the transducer under the spring clip. The adjusting nut knob should
be pointing up and on the side opposite of the spacer bar. Unscrew the knob until it’s at the
stop.
EZ CLAMPING TRANSDUCER TO PIPE
and 'S' Hook
Mate Bit Snap
Chain Extension Sections
Adjusting Nut