3-22
1010PVNFM-3A
Section 3
l
Tighten all TB1 and TB2 terminal block screws.
l
Replace I/O Board and secure with four captive screws.
l
Replace Access Cover and finger tighten captive thumbscrew.
l
Using a flat-blade screwdriver, loosen Terminal Block TB1 and TB2 screws.
l
Connect the wires of the 1012EC Series cable as follows:
3.2.2 1010PVN SUPPLY AND RETURN CONNECTIONS
Terminals for the supply and return sensor connections are located on the Analog Input Mod-
ule as previously shown. For other terminal locations on System 1010N Single Channel and
Dual Channel units, see drawings 1010N-5-7 or 1010N-5D-7, respectively. For terminal loca-
tions on Multi-Channel units, see drawing 1010N-8M-7.
WARNING: Set meter and instrumentation power to OFF when inserting or removing
the module, or when making connections to TB1, TB2, TB3, and TB4.
1. Connect the RTD liquid temperature cable (1012EC) as shown in the table below:
2. Complete the temperature sensor current loop by shorting together terminals 1 and
4 of the unused TB2 temperature sensor terminal block.
3. Ground the voltage sensing leads (terminals 2 and 3 of TB2) by connecting them both
to terminal 5.
1012EC Series
Cable
TERMINAL TB1
WIRE #1 (Black)
to TB1-1
WIRE #2 (White
to TB1-2
WIRE #3 (Green)
to TB1-3
WIRE #4 (Red)
to TB1-4
WIRE #5 GND/SHLD (Blue)
to TB1-5