2-1
1010PVNFM-3A
Section 2
2.
THE 1010PVN INSTALLATION MENU
Programming System 1010PVN requires no special experience or training. This field manual
contains all the necessary information. If you intend to connect this instrument to an external
device, then please have the instruction manual for the device available for reference. The
following paragraphs present a generic menu reference that applies to all configurations of
1010 NEMA meters.
This section contains a general description of how to enter site setup data via the 1010PVN
Installation Menu. It assumes that you have already selected a meter type as described in the
Getting Started section. For convenience, it presents the Installation Menu in the same se-
quence as it appears on the menu screen. Please note, however, that it is not necessary to
program the meter sequentially.
The conventions used with these instructions are:
l
When the text refers to a keypad key, it will be enclosed in “less than” (<) and “greater than”
(>) symbols: e.g., <MENU>, <ENT>, <Up Arrow>, etc. Where a visual of the key is shown;
this means press this key.
l
When the text refers to a menu or menu cell name, it will be enclosed with square brack-
ets: e.g., [Pipe Data], [Channel Enable], etc.
l
Each menu includes an image of its main screen and a diagram of its structure. The rightmost
column of the structure diagram lists the option list choices of the menu cell, if applicable.
General Installation Menu Notes:
l
If a power failure occurs while you are entering or editing data, the entered data may not be
retained in Active Memory.
l
Although you can operate the meter immediately after completion of the site setup, we
recommend that you preserve your settings by saving them under a site name. Site data
can be saved at any time before invoking either the [Recall Site Setup] or the
[Create/Rename Site] commands. When issued, these commands overwrite all data present
in Active Memory.
l
We do not recommend that you attempt to operate the flowmeter at a new site by recalling
and then editing an existing site setup. Each site must have its own set of transducer
install parameters, even if the data from the recalled site setup is identical. Always issue
the [Create/Rename Site] to begin a new site setup. This will fill all the menu cells with
defaults to eliminate the possibility of retaining unwanted parameters.
l
The meter allows you to create your own personal site setup defaults. After creating a site
setup and editing default parameters as desired, save the site using the name [FASTSTRT].
The next time you issue the Create/Name Site command, your custom parameters will
become the system’s defaults.