2-12
1010PVNFM-3A
Section 2
2.2.9 Mod of Elast PSI
Use this menu cell to set the Modular Elasticity.
To enter the value of Modulus of Elasticity:
To enable numeric entry press
Use the number keys to type the Modular Elasticity PSI value.
To register the data press
2.3
THE APPLICATION DATA MENU
Use of the Application Data menu is optional.
CAUTION: An erroneous Viscosity entry could affect the meter’s intrinsic calibration.
Do not alter the default viscosity value of your liquid unless you are sure
of your data.
Use the [Application Data] menu to edit default settings to match your application. When
you specify a Liquid Class, the meter will adjust its operation to accommodate the liquid’s
estimated sonic velocity, viscosity and specific gravity. If necessary, you can edit each liquid
class parameter individually to obtain a closer match with the liquid. The default liquid is
Water at 68
°
F (20
°
C).
The [Temperature Range] menu cell allows you to specify the expected temperature range at
the transducer mounting location. The default setting: [-40
0
F to 250
0
F], matches the standard
991 and 1011 series of transducers but not the 1011H. For higher pipe temperatures, selecting
the proper temperature range allows the flow computer to recommend the appropriate trans-
ducers.
The default pipe configuration (Fully Developed) wll cause the flowmeter to use the conven-
tional Reynolds Compensation Table when compensating for liquid flow profile behavior. The
number of diameters between the upstream configurations and the transducer installation
can be numerically entered via the [Anomaly Diams] menu cell.
ENT
Liquid Class
Water 20C/68F
Liquid Table
Temperature Range
-40F to 250F
Pipe Configuration
Fully Developed
Anomaly Diams
10
Controlotron
Dual Path
SITE 1
Select Liquid Class from Liquid
Application Data
Use this menu cell
to specify the
temperature range
of the liquid.
Use this menu cell to
select an upstream
pipe compensation
correction table.
Use this menu to select