Instrument Calibration
48
515 HC04 - 8 June 2007
Test Menu
The Test menu enables you to view the inputs and outputs to and from the
instrument.
In Calibration Set mode, (by entering the system password) you can control
the outputs and the alarms as described in the table below.
RESET
ACCUM
The Reset Accumulated Totals function clears all of the accumulated
totals and the non-accumulated totals.
Press
or
to select YES, then press the
key. The instrument
makes three beeps to confirm the reset command.
DISPL
TAGS
The Display Tags option determines whether the instrument displays the
default display tags or the user-defined tags. The display tag setting also
defines whether the instrument displays the default error and warning
messages, or the user-defined messages.
Note:
The user-defined tags can be entered into the instrument only by
the manufacturer or the distributor.
Press
or
to select the Display Tags option as follows:
•
DEFAULT
- the instrument displays the default (English) tags
•
USER
- the instrument displays the user-defined tags.
BACK-L
T-OUT
If the backlight timeout is enabled, and there is no user activity (any keys
pressed) for a period of 10 seconds, the display backlight switches off to
save power. The backlight switches on when a key is pressed. Select the
backlight timeout mode as required.
Press
or
to select ENABLE or DISABLE.
RATES
DP
This parameter sets the maximum number of decimal places for
displaying or printing main menu rates.
TOTALS
DP
This parameter sets the maximum number of decimal places for
displaying or printing main menu totals.
↓
→
UNITS PARAMS INPUTS OUTPUTS ALARMS COMMS TM/LOG
SETUP
TEST END
SET
SET
↓
→
UNITS PARAMS INPUTS OUTPUTS ALARMS COMMS TM/LOG SETUP
TEST
END
AINP
n
units
The units are displayed according to the calibration setup for the analog
input. If unused or set to Default the input is 4-20mA and displayed in
mA.
SET
Содержание 515
Страница 4: ...iv 515 HC04 8 June 2007 ...
Страница 8: ...viii 515 HC04 8 June 2007 ...
Страница 16: ...Introduction 8 515 HC04 8 June 2007 ...
Страница 60: ...Instrument Calibration 52 515 HC04 8 June 2007 ...
Страница 80: ...Communications 72 515 HC04 8 June 2007 ...
Страница 82: ...74 515 HC04 8 June 2007 ...
Страница 88: ...80 515 HC04 8 June 2007 ...