Installation
515 HC04 - 8 June 2007
13
Inputs
Analog Input Connections
All analog inputs can accept DC signals ranging from 0-5V, 1-5V and
current signals from 4 to 20mA.
Analog Inputs 1 and 2 (AINP1 / AINP2) can also accept an RTD input
(PT100 or PT500) as well as the standard 0-5V, 1-5V and 4 to 20mA input.
CAUTION
Applying levels of input current above the absolute maximum
rating (100mA) may cause permanent damage to the input
circuitry.
0-5 and 1-5 Volt Inputs
For externally powered voltage transmitters, connect each transmitter to a
pair of input terminals as shown in Figure 4. Refer to
on page 12 for specific terminal numbers for this application.
Figure 4 Externally Powered Voltage Transmitter
Connect internally powered voltage transmitters as shown in Figure 5.
Figure 5 Internally Powered Voltage Transmitter
+
-
-
+
Signal
Transmitter
18
Shield
Analog
Input
(AINPn)
15
-
-
+
+
16
Vo +8...24V DC
DC ground
Signal
Transmitter
18
Shield
Analog
Input
(AINPn)