Menu item Acoustic signalization & ALARM OUT
Submenu for switch OFF the audio indication. This item appears
only, if user allowed in Common parameters Alarm
Confirmation via menu. After entering actual states of audio
indication and ALARM OUT output are displayed. If in active
state, it is possible to deactivate by pressing ENTER key. New
activation can be caused by creation of new alarm or by ending
of alarm and new alarm creation, which caused this action. In
case password requirement function is activated in SW, it is
necessary to enter password first. Way of entering PIN code and
other options are specified in
Menu item Service
Submenu enabling to display value of some Service parameters
of data logger.
Service
–
display of self test of internal voltages
Self test of data logger internal voltage. First value indicates
approximate power voltage (9 to 30 V, see technical
parameters). Second value is voltage of negative source (-14V
to -16V) and third value is voltage of internal back-up battery
(2,6V to 3,3 V).
Service
–
display of firmware version and uP speed
ENTER
MENU
Acoustic sign.
&
ALARM OUT
>>>
ENTER
MENU
Service
>>>
ENTER
MENU
Selftest:
24V -15V 3.0V
ENTER
MENU
Firmware ver.:
5.2.1 6MHz