15/11/2018
Page 16
Excessive power or heating
Cooling soft gels aids in comb
removal.
Reduce voltage. See
electrophoresis instructions.
Slanted lanes (bands)
Gel not fully solidified
Comb warped or at an angle
Gel to solidify for at least 30-
45min.
Check alignment of comb.
runVIEW Issues
Problem
Cause
Solution
No Display / lights
No AC power.
Check if the power supply is unplugged, or if
the AC power source is a problem.
AC power cord is not
connected.
Check AC power cord connections at both
ends. Use the correct cords.
The fuse has blown.
Replace the fuse
Operation stops
Electrophoresis leads are
not connected to the
power supply or the
electrophoresis unit; or
the circuit is broken in
the electrophoresis
system.
Check the connections to the power supply
and within electrophoresis system to make
sure the connection is intact; check the
electrodes to make sure they are intact.
Close the circuit by reconnecting the cables.
Press
START/STOP
to restart the run.
High resistance due to
tape left on a pre-cast
gel, incorrect buffer
concentration, or
insufficient buffer
volumes in the
electrophoresis system.
Make sure that the tape is removed from the
pre-cast gel, that the buffers are prepared
correctly, and the recommended volume of
buffer is added to the electrophoresis unit
and is covering the gel.
Error message
Over voltage (170V
safety limit reached or
exceeded).
Press
START/STOP
button to clear the error
message. Contact Cleaver Scientific’s service
department if the problem persists.
Message
No load is detected.
(1) Check the connections.
(2) Check the buffer condition / buffer
Level.
Alarm message
Maximum power output
reached (30 W).
Warning message for reference.