Page 7
Operating Instructions
Overview
runVIEW is an innovative system designed for real-time size fractionation and
recovery of nucleic acids. runVIEW can maximise the efficiency of DNA
recovery from EtBr and SYBR stained gels by minimising the number of steps
involved in post-electrophoretic purification. runVIEW consists of a multiSUB
™
MSCHOICE system with bluVIEW lid, containing a spectral emission filter and
extractor fan within its viewing pane, and a base unit with integrated power
supply and blue LED gel illuminator.
Installation
Place the runVIEW on a sturdy and level, dry surface. Plug the power cord
into the back of the unit and mains power. The system is now ready for use.
Setting up the Horizontal Gel Tank
Fitting Electrodes
Note the position of the lid on the unit. This shows the correct polarity
and the correct orientation of the cables, black is negative and red
positive.
Remove the lid from the unit. Note if the lid is not removed, fitting the
cables may result in un-tightening of the gold plug and damage to the
electrode.
Screw the cables into the tapped holes as fully as possible so that there
is no gap between the lid and the leading edge of the cable fitting.
Refit the lid.
Fitting Loading Guides
These can be fitted to enhance visibility of the wells if desired. They can be
fitted to the white vinyl platform sheet or to the unit itself.
1. Seat the tray in the unit and note the position of the comb grooves. The
samples run black to red, but the trays can be used frontward or
backwards so ensure that the comb grooves closest to the black
electrode are marked.
2. Remove the tray.