3. PORTASPEC INITIAL SETUP
3-3
3.4.
Power Down and Disassembly
If the Portaspec is to be disassembled for any reason, the following sequence should
be followed.
1.
Turn off x-rays and remove the key from x-ray on key switch.
2.
Exit the Portaspec Control Program
3.
Turn the computer off.
4.
Turn the Portaspec off.
5.
Disconnect power cords from their power sources.
6.
Disconnect Vacuum and water hoses.
3.5.
Transporting
When transporting the Portaspec, especially into areas where the temperature is
below the freezing point, it is necessary to first remove the water from the Portaspec
water lines. Residual water can be forced out with compressed air. If water is
allowed to remain in the lines, it could leak out during transit and contaminate or ruin
electrical components. Beyond that, during cold weather, freezing of the retained
water could cause severe damage to the x-ray tube housing and possibly the x-ray
tube.
The x-ray tube in the Portaspec has a glass envelope. A severe shock during
shipping could crack or break the envelope resulting in a need for an expensive
replacement. Good judgment in selecting proper packing material and carrier is
important.
The Portaspec was shipped from the factory in a carton with sufficient packing to
withstand a reasonable amount of mishandling. If possible, it is desirable to save the
shipping cartons and pallet for future transporting of the Portaspec. However, if the
carton was not retained, a carton made from double wall cardboard with a minimum
burst test of 350 psi and gross weight of 120 lbs. would safely transport the
Power On