2. X-RAY SAFETY
2-4
2.6.1 X-Ray On Interlock Key Switch
.
Part of interlock circuit number 1, the key can only be removed from the switch in the
“Off” position. It is not possible to lock x-rays on. This switch is common to all
Portaspec models.
2.6.2 Rear Panel Safety Interlock Switch
.
This switch is in series with the key switch in interlock circuit number 1. It will not
allow x-rays to be generated without the rear panel properly installed. This switch is
common to all Portaspec models.
Model 2501XBT
Image 1-1
X-Ray On Key Switch
Image 2-6
Rear Panel open
Image 2-7
Rear Panel
Interlock Switches