8.20 TIME-Time Setting
To set or change the system time. Place rotary switch at position “TIME” mode. The time and
date on the third display panel should blink. Enter from left to right in the following format
HH:MM:SS. Press On ( ), under ( ) button printed on the overlay in white word to set the
system date. Press the “ENTER” button to confirm. Exit this function by pressing ESC button.
Except these 2 positions mentioned above, all other position are for system setting, the indica-
tor “SET” shows at left side of LCD when place the rotary switch at any of these system setting
positions. And 2nd functions of buttons are active.
HH: Interval Hour (0~23)
MM: Interval Minute (0~59)
SS: Interval Second (0~59)
MAX: 23:59:59
MIN:00:00:01
8.21 DATE-Date Setting
To set or change the system data. Place rotary switch at position “Date” mode. The time and
date on the third display panel should blink. Enter from left to right in the following format YY:
MM: DD. Press On ( ), under ( ) button printed on the overlay in white word to set the sys-
tem date. Press the “ENTER” button to confirm. Exit this function by pressing ESC button.
Except these 2 positions mentioned above, all other position are for system setting, the indica-
tor “SET” shows at left side of LCD when place the rotary switch at any of these system setting
positions. And 2nd functions of buttons are active.
Y: Interval year (20 "00" ~20 "99")
M: Interval month (1~12)
D: Interval date (1~31)
MAX: 99:12:31
MIN: 00:01:01
9. Maintenance
Periodically wipe the case with a damp cloth, detergent and do not use abrasives or solvents.
WARNING:
To avoid possible electrical shock, disconnect the
thermocouple connectors from the thermometer
before removing the cover.
18
R/S-TYPE
±(0.05% rdg + 2°C) on 0°C to 1767°C
±(0.05% rdg + 4°F) on 32°F to 3212°F
CHA (RTD) Measuring Range:
Acceptable RTD Type:
Pt385(100ohm)
、
Pt3916(100ohm)
、
Pt3926(100ohm).
Temperature:
-200°C to 800°C, -328°F to 1472°F
Resolution:
0.01, 0.1 for above ±200 °C/°F.
Accuracy:
±0.05% rdg + 0.2°C or ±0.05% rdg + 0.4°F
Resistance:
0 ohm to 199.99 ohm, 200.0 ohm to 400.0 ohm.
Resolution:
0.01, 0.1 for above 200 ohm.
Accuracy:
±0.05% rdg + 0.1 ohm.
CHB (ADAPTER) Measuring Range:
Accuracy:
±(0.1% rdg + 10dgts) on 200mV
±(0.5% rdg + 4°dgts) on 2000mV
Stated accuracy at 23°C ± 5°C, <75% R.H.
Temperature Coefficient:
0.1 times the applicable accuracy specification per °C from 0°C to
18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection:
24V DC or 24V AC rms maximum input voltage on any combination of
input pins.