Top Side:   

1.

 

Socket of power (DC 12V, more then 100mA). 

2.

 

RS232 (Optical Interfaced) port. 

3.

 

Socket of CHA (channel A) for RTD input. 

4.

 

Sockets of K type thermocouples T1 and T2. 

 

Connection Of RTD 

 

Configuration of Mini-Din Connector 

 

Wire connection to RTD 

LCD: 

 

Number Display 

Function 

1 SET 

Setting 

mode 

Pt385/3916/3926  Type of RTD at CHA (channel A)

 

 

Low battery 

APO 

Auto power off 

HOLD 

Hold measured reading 

 

 

8.17 APO-Auto Power Off Time Setting 

Use this function to change the time for the auto power off (APO) function. Place rotary 

switch at position “INVT”, and the indicator “APO” and the main display will blink on the LCD, 
and the previous setting will be displayed. The default time for APO is 1 minute. Press On ( ), 
under ( ) button printed on the overlay in white word to set the APO time. On ( ), under ( ) 
direction key: Time value, right ( ), left ( ) direction key: Time figure. Press “CANCEL” button 
to back to previous setting. Confirm selection and exit setting mode by pressing “confirm (

)” 

button. Press “ESC” button to abort change and exit the setting mode with previous setting re-
mains unchanged. Confirm selection and exit setting mode by pressing “confirm (

)” button. 

MAX: 9998 minutes 

MIN: 0001 minutes 
Disable APO function: To set the time at 9999 will disable APO function. 
APO Function will be Disabled also in Follow situation: 
External power is used. 
RS-232 function is used. 
MAX, MIN, MAX-MIN, AVG function is used. 
Data logger function is used.

 

 
8.18 INVT- Set Time interval of data logging 

Place rotary switch at position “INVT” mode to set the time. The indicator “INV” will be blink 

on the top-right of the LCD and the pervious settings are displayed. Press On ( ), under ( ) 
button printed on the overlay in white word to change the time interval. Setting is from left to 
right of the following format (HH:MM:SS). 

Pressing overlay “ENTER” button to confirm. To exit this function, press the ESC button. 

Except these 2 positions mentioned above, all other position are for system setting, the indicator 
“SET” shows at left side of LCD when place the rotary switch at any of these system setting po-
sitions. And 2nd functions of buttons are active. 
HH: Interval Hour (0~23) 
MM: Interval Minute (0~59) 
SS: Interval Second (0~59) 
MAX: 23:59:59 
MIN: 00:00:01

 

 
8.19 PERIOD- Set length of Time of data logging 

Place the rotary switch at position “Period” to set the total length of time for data recording in 

the format of DD.HH.MM. At this position, indicator “PER” blinks at right upper corner of the 
LCD and the third display (the lower right display) shows the previous setting with the HH 
blinks. The blinking number can be changed with upward ( ) or downward ( ) direction but-
tons, and blinking position can be changed with right ( ) or left ( ) direction buttons. 

17 

Содержание 506A

Страница 1: ...OPERATOR S MANUAL MODEL 506A...

Страница 2: ...r ID 001 Ex PC send 001L00006 r then meter response 001L0000111554170704111220420900115000001 r PC sends 001D00006 r then meter will dump all of data of group 6th 001D00006 rxxxxxx rxxxxxx rxxxxxx r M...

Страница 3: ...of Buttons 9 8 1 D1 Setting the Main Display 9 8 2 D2 Setting Second Display 10 8 3 D3 Setting Third Display 10 8 4 T1 TYPE and T2 TYPE Setting the Types of Thermocouple for T1 or T2 10 8 5 T1 ADJ an...

Страница 4: ...s 5 Alarm Counter AC 5 S Counts 5 Saved Data SD mode 1 mode 2 34 S GroupID SC c f mode T1_type T2_type Time Value r 5 5 1 1 1 1 12 6 2 Saved Data SD mode 3 34 S GroupID SC c f dp cha_type Time Value r...

Страница 5: ...trument contains no operator serviceable part Service should be done by qualified person only Do not use the meter if the meter or test leads look damaged or if you suspect that the meter is not opera...

Страница 6: ...S TYPE Resolution 1 0 C to 1767 C 32 F to 3212 F N TYPE Resolution 0 1 50 C to 1300 C 58 F to 2372 F Resolution 0 1 C 1 C 0 1 F 1 F Auto Ranging 2 2 Accuracy K J T E TYPE 0 05 rdg 0 3 C on 50 C to 13...

Страница 7: ...the indica tor SET shows at left side of LCD when place the rotary switch at any of these system setting positions And 2nd functions of buttons are active Y Interval year 20 00 20 99 M Interval month...

Страница 8: ...function APO Function will be Disabled also in Follow situation External power is used RS 232 function is used MAX MIN MAX MIN AVG function is used Data logger function is used 8 18 INVT Set Time inte...

Страница 9: ...e blinking position with and key Press ESC key to end the set up mode indicator C or F ends blinking and indicator SET disappears but keep the previous setting unchanged Press CANCEL key to return to...

Страница 10: ...ory Clear function or to leave this function by turning the Rotary Switch to an other position SET SET function Select the channels for data logging When the blinking indicator shows SET press key to...

Страница 11: ...on marked with turning the rotary switch to any other position automatically turns on Model 506A5 Num bers and marks appear in the LCD display When APO Auto Power Off is enabled refer to 8 17 for enab...

Страница 12: ...refer to section 8 18 To read the logged data please refer to section 8 12 8 8 11 CHB RANGE Select the Resolution for Cannel B CHB Channel B can accept the voltage signal up to 2 volts above 2 volts...

Страница 13: ...And not function when LOG function is set as multi channel logging During measuring press SAVE button to record current measurement reading of the channel shown in Main Display and time To read the r...

Страница 14: ...are known the offset adjustments can be set to each of T1 and T2 for compensating the off set er ror 10 RTD TYPE 8 6 CHA TYPE Selecting the Types of RTD for CHA Besides accepting the RTD of type Pt 38...

Отзывы: