2. Specifications 

Display:

 4 1/2 digit liquid crystal display (LCD) with a maximum reading of 19999. 

Low Battery Indication: 

The “ ” displayed and blinking when the battery voltage drops below 

the operating level. 

Measurement Rate: 

2.5 times per second (nominal). 

Continuous Record: 

1. Single channel recording: 32000 records (T1 or T2 or CHA or CHB). 
2. Multi channel recording: 6500 records (T1+T2 or T1+T2+CHA or T1+T2+CHB).   

ALM Record:

 5400 samples with start/stop time. (Max.)   

Single Record: 

6500 samples with real-time data. (Max.)   

Operating Environment:

 0°C to 50°C at < 70% R.H. 

Storage Temperature:

 -20°C to 60°C, 0 to 80% R.H. with battery removed from meter. 

Safety:

 CE 

Power:

 Single standard 9-volt battery (NEDA 1604, JIS 006P, IEC 6F22). 

Battery Life: 

100 hours typical with carbon-zinc.   

Dimensions:

 192mm (H) x 91mm (W) x 52.5mm (D).   

Weight:

 435g including battery. 

Software download link: 
http://WWW.chy-meter.com//data/outer_web/1492-00506A/1492-00506A-102.zip  

Accessories:

 

K-type thermocouples (2 pcs) 
Co-axial cable (RTD) (1pcs) 
RS232 transmission cable (1pcs) 
9V battery (installed) (1pcs) 
Operating manual 

 

2.1 Electronic Specifications: 
Thermocouple(T1

T2)Measuring Range: 

K-TYPE (Resolution 0.1°): -200°C to 1372°C, -328°F to 2501°F 

J-TYPE (Resolution 0.1°): -210°C to 1200°C, -346°F to 2192°F 
T-TYPE (Resolution 0.1°): -200°C to 400°C, -328°F to 752°F 
E-TYPE (Resolution 0.1°): -210°C to 1000°C, -346°F to 1832°F 
R-TYPE (Resolution 1°): 0°C to 1767°C, 32°F to 3212°F 
S-TYPE (Resolution 1°): 0°C to 1767°C, 32°F to 3212°F 
N-TYPE (Resolution 0.1°): -50°C to 1300°C, -58°F to 2372°F 

Resolution: 0.1°C /1°C, 0.1°F /1°F (Auto Ranging) 

 

2.2 Accuracy: 
K/J/T/E TYPE 

 ±(0.05% rdg + 0.3°C) on -50°C to 1370°C 

±(0.05% rdg + 0.7°C) on -50°C to -210°C 

±(0.05% rdg + 0.6°F) on -58°F to 2501°F 
±(0.05% rdg + 1.4°F) on -58°F to -346°F 

N -TYPE 

 ±(0.05% rdg + 0.8°C) on -50°C to 0°C 
±(0.05% rdg + 0.4°C) on 0°C to 1300°C 
±(0.05% rdg + 1.6°F) on -58°F to 32°F 
±(0.05% rdg + 0.8°F) on 32°F to 2372°F 

A.

 

Command Sets (from Remote to Instrument) 

A. 1 Instrument Setup (%) , only use for broadcasting 

Command Syntax 

Command Name 

Description 

%ID 

I

 

ID

 

New ID 

ID: original ID 

ID

: New ID 

%000 

R

  

Read ID 

 

%ID

 D

 DE 

Setup description DE: 

Descriptive Text 

%ID 

T

 TI 

Timer setup 

TI: timer count 

%ID 

L

 LM 

Limit setting 

LM: Hi/Lo limit and Range 

%ID 

S

 TSEC 

Log sampling time 

TSEC: Log sampling time 

%ID 

M

 

Clean Memory 

 

Unknown or Error Echo

?

 

 

A.2 Instrument Status ($) 

Command Syntax 

Command Name 

Description 

$ID 

G

 

Read group counter   

Echo: ! ID 

G

 SC , LC , AC 

SC: saved group count 
LC: log group count 
AC: alarm count 

$ID 

D

 

Read description 

Echo: ! ID 

D

 DE 

$ID 

T

 

Read Timer 

Echo: ! ID 

T

 TI 

$ID 

L

 

Read Limits Value 

Echo: ! ID 

L

 LMS 

$ID 

S

 

Read Sampling Time 

Echo: ! ID 

S

 TSEC 

$ID 

B

 

Read Battery Voltage 

Echo: ! ID 

B

 VB 

Unknown or Error Echo 

 

A.3 Measurement (#)

 

Command Syntax 

Command Name 

Description 

#ID 

N

 

Read current data   

Echo: > ID 

N

 #IDN 

#IDN: Current Data 

#ID 

[SC] 

Read saved data 

Echo: > ID 

S

 SD 

SD: Saved Data 

#ID 

L

 [LC] 

Read log status   

Echo: > ID 

L

 LS 

#ID 

D

 [LC] 

Read log data 

Echo: > ID 

D

 LD 

#ID G [Start] , [Stop] 
                5    1    5 

Read saved data group 

Echo: > ID 

G

 SDG 

 

#ID B   

Read alarm log data 

Echo: > ID 

B

 AD 

AD: Alarm Log Data 

Unknown or Error Echo

?

 

 

19 

Содержание 506A

Страница 1: ...OPERATOR S MANUAL MODEL 506A...

Страница 2: ...r ID 001 Ex PC send 001L00006 r then meter response 001L0000111554170704111220420900115000001 r PC sends 001D00006 r then meter will dump all of data of group 6th 001D00006 rxxxxxx rxxxxxx rxxxxxx r M...

Страница 3: ...of Buttons 9 8 1 D1 Setting the Main Display 9 8 2 D2 Setting Second Display 10 8 3 D3 Setting Third Display 10 8 4 T1 TYPE and T2 TYPE Setting the Types of Thermocouple for T1 or T2 10 8 5 T1 ADJ an...

Страница 4: ...s 5 Alarm Counter AC 5 S Counts 5 Saved Data SD mode 1 mode 2 34 S GroupID SC c f mode T1_type T2_type Time Value r 5 5 1 1 1 1 12 6 2 Saved Data SD mode 3 34 S GroupID SC c f dp cha_type Time Value r...

Страница 5: ...trument contains no operator serviceable part Service should be done by qualified person only Do not use the meter if the meter or test leads look damaged or if you suspect that the meter is not opera...

Страница 6: ...S TYPE Resolution 1 0 C to 1767 C 32 F to 3212 F N TYPE Resolution 0 1 50 C to 1300 C 58 F to 2372 F Resolution 0 1 C 1 C 0 1 F 1 F Auto Ranging 2 2 Accuracy K J T E TYPE 0 05 rdg 0 3 C on 50 C to 13...

Страница 7: ...the indica tor SET shows at left side of LCD when place the rotary switch at any of these system setting positions And 2nd functions of buttons are active Y Interval year 20 00 20 99 M Interval month...

Страница 8: ...function APO Function will be Disabled also in Follow situation External power is used RS 232 function is used MAX MIN MAX MIN AVG function is used Data logger function is used 8 18 INVT Set Time inte...

Страница 9: ...e blinking position with and key Press ESC key to end the set up mode indicator C or F ends blinking and indicator SET disappears but keep the previous setting unchanged Press CANCEL key to return to...

Страница 10: ...ory Clear function or to leave this function by turning the Rotary Switch to an other position SET SET function Select the channels for data logging When the blinking indicator shows SET press key to...

Страница 11: ...on marked with turning the rotary switch to any other position automatically turns on Model 506A5 Num bers and marks appear in the LCD display When APO Auto Power Off is enabled refer to 8 17 for enab...

Страница 12: ...refer to section 8 18 To read the logged data please refer to section 8 12 8 8 11 CHB RANGE Select the Resolution for Cannel B CHB Channel B can accept the voltage signal up to 2 volts above 2 volts...

Страница 13: ...And not function when LOG function is set as multi channel logging During measuring press SAVE button to record current measurement reading of the channel shown in Main Display and time To read the r...

Страница 14: ...are known the offset adjustments can be set to each of T1 and T2 for compensating the off set er ror 10 RTD TYPE 8 6 CHA TYPE Selecting the Types of RTD for CHA Besides accepting the RTD of type Pt 38...

Отзывы: