2. Specifications
Display:
4 1/2 digit liquid crystal display (LCD) with a maximum reading of 19999.
Low Battery Indication:
The “ ” displayed and blinking when the battery voltage drops below
the operating level.
Measurement Rate:
2.5 times per second (nominal).
Continuous Record:
1. Single channel recording: 32000 records (T1 or T2 or CHA or CHB).
2. Multi channel recording: 6500 records (T1+T2 or T1+T2+CHA or T1+T2+CHB).
ALM Record:
5400 samples with start/stop time. (Max.)
Single Record:
6500 samples with real-time data. (Max.)
Operating Environment:
0°C to 50°C at < 70% R.H.
Storage Temperature:
-20°C to 60°C, 0 to 80% R.H. with battery removed from meter.
Safety:
CE
Power:
Single standard 9-volt battery (NEDA 1604, JIS 006P, IEC 6F22).
Battery Life:
100 hours typical with carbon-zinc.
Dimensions:
192mm (H) x 91mm (W) x 52.5mm (D).
Weight:
435g including battery.
Software download link:
http://WWW.chy-meter.com//data/outer_web/1492-00506A/1492-00506A-102.zip
Accessories:
K-type thermocouples (2 pcs)
Co-axial cable (RTD) (1pcs)
RS232 transmission cable (1pcs)
9V battery (installed) (1pcs)
Operating manual
2.1 Electronic Specifications:
Thermocouple(T1
、
T2)Measuring Range:
K-TYPE (Resolution 0.1°): -200°C to 1372°C, -328°F to 2501°F
J-TYPE (Resolution 0.1°): -210°C to 1200°C, -346°F to 2192°F
T-TYPE (Resolution 0.1°): -200°C to 400°C, -328°F to 752°F
E-TYPE (Resolution 0.1°): -210°C to 1000°C, -346°F to 1832°F
R-TYPE (Resolution 1°): 0°C to 1767°C, 32°F to 3212°F
S-TYPE (Resolution 1°): 0°C to 1767°C, 32°F to 3212°F
N-TYPE (Resolution 0.1°): -50°C to 1300°C, -58°F to 2372°F
Resolution: 0.1°C /1°C, 0.1°F /1°F (Auto Ranging)
2.2 Accuracy:
K/J/T/E TYPE
±(0.05% rdg + 0.3°C) on -50°C to 1370°C
±(0.05% rdg + 0.7°C) on -50°C to -210°C
±(0.05% rdg + 0.6°F) on -58°F to 2501°F
±(0.05% rdg + 1.4°F) on -58°F to -346°F
N -TYPE
±(0.05% rdg + 0.8°C) on -50°C to 0°C
±(0.05% rdg + 0.4°C) on 0°C to 1300°C
±(0.05% rdg + 1.6°F) on -58°F to 32°F
±(0.05% rdg + 0.8°F) on 32°F to 2372°F
2
A.
Command Sets (from Remote to Instrument)
A. 1 Instrument Setup (%) , only use for broadcasting
Command Syntax
Command Name
Description
%ID
I
ID
New ID
ID: original ID
ID
: New ID
%000
R
Read ID
%ID
D
DE
Setup description DE:
Descriptive Text
%ID
T
TI
Timer setup
TI: timer count
%ID
L
LM
Limit setting
LM: Hi/Lo limit and Range
%ID
S
TSEC
Log sampling time
TSEC: Log sampling time
%ID
M
Clean Memory
Unknown or Error Echo
?
A.2 Instrument Status ($)
Command Syntax
Command Name
Description
$ID
G
Read group counter
Echo: ! ID
G
SC , LC , AC
SC: saved group count
LC: log group count
AC: alarm count
$ID
D
Read description
Echo: ! ID
D
DE
$ID
T
Read Timer
Echo: ! ID
T
TI
$ID
L
Read Limits Value
Echo: ! ID
L
LMS
$ID
S
Read Sampling Time
Echo: ! ID
S
TSEC
$ID
B
Read Battery Voltage
Echo: ! ID
B
VB
Unknown or Error Echo
?
A.3 Measurement (#)
Command Syntax
Command Name
Description
#ID
N
Read current data
Echo: > ID
N
#IDN
#IDN: Current Data
#ID
S
[SC]
Read saved data
Echo: > ID
S
SD
SD: Saved Data
#ID
L
[LC]
Read log status
Echo: > ID
L
LS
#ID
D
[LC]
Read log data
Echo: > ID
D
LD
#ID G [Start] , [Stop]
5 1 5
Read saved data group
Echo: > ID
G
SDG
#ID B
Read alarm log data
Echo: > ID
B
AD
AD: Alarm Log Data
Unknown or Error Echo
?
19