Rotary Switch: 

1.

 

Off position. 

2.

 

Work position. 

3.

 

Main/2nd/3rd display setup. 

4.

 

Sockets of K type thermocouples T1 and T2 setup. 

5.

 

RTD (CHA) setup. 

6.

 

ADAPTER (CHB) setup. 

7.

 

Data processing. 

8.

 

Battery test. 

9.

 

Setting. 

 

 

4. Battery Installation 

Remove the two screws from the back of the meter and lift off the battery case. 

 

Note: 
It is need to replace the battery less than 20 sec. to avoid the time of system was reset. 

 

Press direction keys 

 or    to change the indicator among “READ MEM SAVE”, “READ 

MEM LOG” and “READ MEM LIMIT”. 
Press “

” key to enter the indicated set of recorded data. 

 
In “READ MEM SAVE” and “READ MEM LIMIT”, press direction keys 

 or   to move 

among the data with the measurement reading shown in main display, sequence number in sec-
ond display and time in third display. 

 
There may be more than one group of data in “READ MEM LOG”, right after entering 

“READ MEM LOG” data, second show GRP 0001, press direction keys 

 or   to move 

among the different group of data, press “

” key to enter the group indicated in second display. 

Press direction keys   or    to move to the data of different channels in the same record and 
press direction keys 

 or    to move among the different records of same channel.   

 
Press “ESC” button to leave “VIEW” function and return to “Memory Access Functions Se-

lection mode” with blinking indicator “SET” at upper-right corner.   

 
 Press direction keys   or    to change the blinking indicator to “SET” (SET function) or 

“CLR” (Memory Clear function), or to leave this function by turning the Rotary Switch to an-
other position. 

 
SET (SET function : Select the channels for data logging) 
 
When the blinking indicator shows “SET”, press “

” key to enter SET function. The indica-

tor “SET” stops blinking and another small indicator “SET” appears at left side of LCD. 

 
Note: When there is data in the memory, you can not enter “SET” function.   

       To erase recorded data, please execute “CLR” function mention above. 

 
Right after entering “SET function”, previously selected channel is shown at top of LCD. 
Press direction keys 

 or    to change the channel among “T1”, “T2”, “CHA”, “CHB”, 

“T1T2”, “CHA T1T2” and “CHB T1T2”. 
 
Press “

” key to confirm selection and return to function selection with indicator “SET” 

blinking at upper-right corner of LCD. 

 
Press direction keys 

 or    to change the blinking indicator to “CLR” (Memory Clear 

function) or “VIE” (VIEW function), or to leave this function by turning the Rotary Switch to 
another position. 

 
Note: When single channel (i.e. T1 or T2 or CHA or CHB) is selected, the selected channel 

will be automatically assigned to Main Display and overwrite the previous setting of Main Dis-
play. When multi channel (i.e. “T1 T2” or “CHA T1 T2” or “CHB T1 T2”) is selected the first 
channel (i.e. T1 for “T1 T2” or “CHA” for “CHA T1 T2” or CHB for “CHB T1 T2”) will be au-
tomatically assigned to Main Display and overwrite the previous setting of Main Display.   
If you do not want this automatic assigned channel shown in Main Display, please re-set the 
Main Display (see section 8.1 for setting Main Display). This resetting of Main Display does not 
influence the set channel(s) for data logger. 

15 

Содержание 506A

Страница 1: ...OPERATOR S MANUAL MODEL 506A...

Страница 2: ...r ID 001 Ex PC send 001L00006 r then meter response 001L0000111554170704111220420900115000001 r PC sends 001D00006 r then meter will dump all of data of group 6th 001D00006 rxxxxxx rxxxxxx rxxxxxx r M...

Страница 3: ...of Buttons 9 8 1 D1 Setting the Main Display 9 8 2 D2 Setting Second Display 10 8 3 D3 Setting Third Display 10 8 4 T1 TYPE and T2 TYPE Setting the Types of Thermocouple for T1 or T2 10 8 5 T1 ADJ an...

Страница 4: ...s 5 Alarm Counter AC 5 S Counts 5 Saved Data SD mode 1 mode 2 34 S GroupID SC c f mode T1_type T2_type Time Value r 5 5 1 1 1 1 12 6 2 Saved Data SD mode 3 34 S GroupID SC c f dp cha_type Time Value r...

Страница 5: ...trument contains no operator serviceable part Service should be done by qualified person only Do not use the meter if the meter or test leads look damaged or if you suspect that the meter is not opera...

Страница 6: ...S TYPE Resolution 1 0 C to 1767 C 32 F to 3212 F N TYPE Resolution 0 1 50 C to 1300 C 58 F to 2372 F Resolution 0 1 C 1 C 0 1 F 1 F Auto Ranging 2 2 Accuracy K J T E TYPE 0 05 rdg 0 3 C on 50 C to 13...

Страница 7: ...the indica tor SET shows at left side of LCD when place the rotary switch at any of these system setting positions And 2nd functions of buttons are active Y Interval year 20 00 20 99 M Interval month...

Страница 8: ...function APO Function will be Disabled also in Follow situation External power is used RS 232 function is used MAX MIN MAX MIN AVG function is used Data logger function is used 8 18 INVT Set Time inte...

Страница 9: ...e blinking position with and key Press ESC key to end the set up mode indicator C or F ends blinking and indicator SET disappears but keep the previous setting unchanged Press CANCEL key to return to...

Страница 10: ...ory Clear function or to leave this function by turning the Rotary Switch to an other position SET SET function Select the channels for data logging When the blinking indicator shows SET press key to...

Страница 11: ...on marked with turning the rotary switch to any other position automatically turns on Model 506A5 Num bers and marks appear in the LCD display When APO Auto Power Off is enabled refer to 8 17 for enab...

Страница 12: ...refer to section 8 18 To read the logged data please refer to section 8 12 8 8 11 CHB RANGE Select the Resolution for Cannel B CHB Channel B can accept the voltage signal up to 2 volts above 2 volts...

Страница 13: ...And not function when LOG function is set as multi channel logging During measuring press SAVE button to record current measurement reading of the channel shown in Main Display and time To read the r...

Страница 14: ...are known the offset adjustments can be set to each of T1 and T2 for compensating the off set er ror 10 RTD TYPE 8 6 CHA TYPE Selecting the Types of RTD for CHA Besides accepting the RTD of type Pt 38...