8.2 D2-Setting Second Display (Lower-Left Display)
Second display can be set to show the reading of T1 or T2. When the power is turn on, the me-
ter remains in the previous setting. Turn the rotary switch to the position “D2”, and use right ( ) or
left ( ) direction buttons to change between T1 and T2.
8.3 D3-Setting Third Display (Lower-Right Display)
Third display can be set to show the value of T1-T2, Date (YY: MM: DD) and time
(HH:MM:SS). When the power is turn on, the meter remains in the previous setting. Turn the ro-
tary switch to the position “D3”, and use right ( ) or left ( ) direction buttons to change between
T1 and T2.
8.4 T1 TYPE and T2 TYPE-Selecting the Types of Thermocouple for T1 or T2
Both T1 and T2 can accept the thermocouples of type K, J, T, E, R, S and N. When the power is
turn on, the meter remains in the previous setting.
Turn the rotary switch to the position “T1 Type” or “T2 Type”, current selection is shown at up-
per-right corner of LCD, adjacent to unit of temperature, change type of thermo couple with
upward ( ) or downward ( ) direction buttons after right type is shown, and press confirm key
(
) to select.
8.5 T1 ADJ and T2 ADJ-Setting the Off-Set Adjustment for T1 or T2
If the offset error of your thermo couples, and to which socket those particular thermocouples
are known, the offset adjustments can be set to each of T1 and T2 for compensating the off-set er-
ror.
10
RTD TYPE
8.6 CHA TYPE-Selecting the Types of RTD for CHA
Besides accepting the RTD of type Pt 385, Pt3916 and Pt3926, this meter can also display the
resistance value in ohm ( ) of the sensor.
When the power is turn on, the meter remains in the previous setting. Turn the rotary switch to the
position “CHA-Type”, current selection is shown at left side of LCD for RTD sensors and right
hand side, for ohm. Change the types with upward ( ) or downward ( ) direction buttons after
right type is shown, press return key (
) to select.
8.7 CHA ADJ-Setting the Off-Set Adjustment for RTD
If the off-set error of your RTD is known, the off-set adjustments to the channel A (CHA).
Place the rotary switch at positions “ADJ” just below “CHA-Type”.
Select the digit you want to adjust with right ( ) or left ( ) direction buttons, (Setting of the ad-
justment value is done one individual digit after the other).
Change the value of blinking digit with upward ( ) or downward ( ) direction buttons. After
reach desired value, press return key (
) to confirm.
8.8 CHA RANGE-Select the Resolution for Channel A (CHA)
Place the rotary switch at position “CHA RANGE”, change among Auto (“--.-.-” and “AUT” is
shown in display) and “0.1” (“---.-” and “MAN” is shown in display) and “0.01” (“--.--” and
“MAN” is shown in display).
When “Auto” is selected, the meter will select most suitable resolution between “0.1” and
“0.01”, while the other two, the reading will be fixed in the selected resolution.