8.9 CHB TYPE-Select the Unit for Channel B
CHB only shows the voltage, and the unit represented by this voltage has to be given. At posi-
tion “CHB TYPE”, change the unit with upward ( ) or downward ( ) direction buttons among
AAA,
→˚
C
→
ppt
→μ
s
→
ms
→
s
→
FPM
→
%RH
→
%
→
ppm
→
mV
→
V
→μ
A
→
mA
→
A
→
ohm.
The first unit, AAA, can be redefined by user for other special unit, see next section. Press the
confirm button (
) to select the unit. The selection is kept even power turned off.
CHB is for a transducer with output of 1 mV = 1 “unit”.
CHB always measures the input voltage with mV.
The “unit” is only a human readable indicator for the convenience of user, it does not change
the measurement.
The “unit” can be set as “mV” or “V” or “%RH” or “%” or “ohm” or “ABC”, and it only make
sense when it is set according to the output property of the transducer.
For example if we connect a humidity transducer with output 1 mV = 1 %RH to the CHB
(channel B) of the meter, the “unit” should be set as “%RH”. If user does not do so, then the
the unit becomes meaningless, but the measurement will not influenced.
8.10 CHB PARA-Create Special Unit for Channel B (CHB)
Changing each individual character of “AAA” to one of following can create special units.
(Characters available: upper case characters A to Z, lower case a to z, o, /, %,
μ
, dot and blank
space).
Place rotary switch at position “CHB PARA”, This user defined unit (set as “AAA” when
shipment) shows at right-upper corner of LCD with the left character blinking. The blinking char-
acter can be changed with upward ( ) or downward ( ) direction buttons.
With right ( ) or left ( ) direction buttons, the blinking can be moved to another character. Af-
ter all 3 character are changed to the desired ones, press confirm button (
).
(AUTO) (MANUAL)
12
Main Display
7.7 SAVE :
SAVE: This function only apply to the channel shown in main display. And not function when
LOG function is set as multi-channel logging.)
During measuring, press “SAVE” button to record current measurement reading of the channel
shown in Main Display and time.
To read the recorded data, please refer to section 8.12
To set the channel for Main Display please refer to section 8.1
To set the channel for LOG function please refer to section 8.12
7.8 ALMLOG: (This function only apply to the channel shown in main display. And not
function when LOG function is set as multi-channel logging)
Every time when the measurement reading of the channel shown in Main Display the Alarm
Log function records the first measurement reading which is out of limits and time, when the
measurement value returns to within the limits the last out of limit data and time will also be rec-
orded.
Press the “ALMLOG” button, to activate the alarm log function, and the indicators “LIMIT” and
“LOG” are shown at lower-right corner of the LCD.
Press the “ALMLOG” button again to disable this function.
To read the recorded data, please refer to section 8.12
To set the channel for Main Display please refer to section 8.1
To set the channel for LOG function please refer to section 8.12
8. System Setting and second Functions of Buttons:
Except these 3 positions mentioned above, all other position are for system setting, the indicator
“SET” shows at left side of LCD when place the rotary switch at any of these system setting posi-
tions. And 2
nd
functions of buttons are active.
8.1 D1-Setting the Main Display
Main display can be set to show the reading of any of the 4 channels. Place rotary switch at po-
sition “D1”, and select the channel with “right ( )”or “left ( )” direction keys to change the chan-
nel. “D1”, name of active channel is shown at top of LCD. Press “CANCEL” button to back to
previous setting. Press “ESC” button to abort change and exit the setting mode with previous set-
ting remains unchanged. Confirm selection and exit setting mode by pressing “confirm (
)”
button.