Manual Operation
3.3.7.5.2 SETUP
1
2
3
4
5
6
+12VA APIGND
V SET
I SET
V
MEAS.
I
MEAS.
+
+
−
−
DC Power
Supply
DVM1
DVM2
−
+
Figure 3-100
3.3.7.5.3 Calibration Procedure (Example: Model 62012P-80-60)
(1) In CALIBRATION page, press “
4
” or turn “Rotary” (
) knob to set CHOICE = 4.
(2) Press “
ENTER
” to confirm entering into APG Voltage Calibration options as Figure 3-101
shows.
[ A P G C U R R E N T C A L I B R A T I O N ]
C H E C K A P G C O N N E C T I O N A N D P R E S S [ E N T E R ]
( S E T ) I N P U T V O L T A G E F O R S E T T I N G = 0 . 5 V
A C T U A L A P G
I N P U T V O L T A G E
= 0 . 6 6 V
( S E T ) I N P U T V O L T A G E F O R S E T T I N G = 8 . 0 V
A C T U A L A P G
I N P U T V O L T A G E
= 8 . 6 8 V
Figure 3-101
NOTICE
When entering into the CALILBRATION page, be sure to check the interface connection on
the rear panel is correct and then press “
ENTER
” to start calibration.
(3) When in the APG Current Calibration pages and the connection is correct, press “
ENTER
”
to confirm.
(4) It will ask users to input about 0.5V voltage signal (Pin 4), and press “
ENTER
” the cursor
will stop at position [1] as Figure 3-102 shows. Adjust the Power Supply to 0.5V±0.2V
and use DVM1 to measure the reading of Power Supply. Input the voltage (2 digits after
decimal point) read by DVM 1 to position [1] and press “
ENTER
” to confirm.
3-61