Programmable DC Power Supply 62000P Series Operating & Programming Manual
2. Use the numeric (
0
~
1
) keys or “Rotary”(
) knob to select “ON” or “OFF” mode.
3. Press
“
ENTER
” to confirm.
4. Press
“
EXIT
” to return to MAIN PAGE.
NOTICE
1. BUZZER has two options: ON / OFF.
2. When the BUZZER is set to ON, press any key or turn the rotary knob will beep once to
remind user.
3. When the BUZZER is set to ON and the BUZZER will beep continuously if system
protection occurs to remind user.
4. When BUZZER is set to OFF then it will not beep in any situation.
3.3.1.6 POWER
ON
STATUS
It loads the default output status automatically after powered on, so that users don’t have to
set it again.
1. Use
“
”, “
” keys to move the cursor to the column to be set.
[ S Y S T E M S E
T
U
P ]
G
P I B A
D
D R =
1
R S - 2 3 2 B A U D R A T E =
9 6 0 0
A P G
=
V
V r e f ( V ) =5
B
A C
K
L I G
H T =
H I G H
B
U Z Z
E
R =
O N
P O W E R O N S T A T U S = D E F A U L T
Figure 3-13
2. Press
“
0
” or “
2
” to set the value, or use ““Rotary” (
) knob to set it.
POWER ON STATUS has three options: DEFAULT / LAST TURN OFF STATUS /
USER DEFINITION.
When it is set to DEFAULT, it means the output state is not defined specifically, which
is V = 0.00V ; I = 0.00A ; OUTPUT = OFF.
When it is set to LAST TURN OFF STATUS, the instrument will log the command
voltage, command current and output state before powering it off, so that the state before
powered off can be obtained for next power-on.
3-10