Manual Operation
13. Press “
EXIT
” to return to MAIN PAGE.
Improper shunt range selection may cause damage to the current shunt.
NOTICE
The calibration point may be different for other models (non 62012P-80-60), please operate it
following the instructions displayed.
3.3.7.3 Current Output (PROG.) Calibration
3.3.7.3.1 Hardware Requirements
Device
Suggest Model or Capacity
DVM
HP 34401A or equivalent DVM
CURRENT SHUNT
Prodigit 7530 or equivalent
Table 3-10
3.3.7.3.2 SETUP
62000P
DC Power
Supply
+
v
−
v
+
−
DVM
+
sv
−
sv
Current Shunt
Figure 3-85
3.3.7.3.3 Calibration Procedure (Example: Model 62012P-80-60)
(1) In CALIBRATION page, press “
3
” or turn “Rotary” (
) knob to set CHOICE=3.
(2) Press “
ENTER
to confirm and entering into current calibration options as Figure 3-86
shows.
3-53