Programmable DC Power Supply 62000P Series Operating & Programming Manual
[ P R O T E C T I O N ]
O V P
O =
8 8 . 0 V
Σ
O C P
O =
6 3 . 0 A
Σ
O P P
O = 1 2 6 0 . 0 W
F O L D B A C K F = D I S A B L E
Figure 3-45
3.3.3.6 Setting Procedure for APG in Series or Parallel
Operation in series and parallel can also be applied via APG. See section 3.3.1.3 and 3.3.3.1
to 3.3.3.6 for detail information.
3.3.3.6.1 Series Setting
To connect 5 62012P-80-60 DC Power Supplies in series for operation and set the APG
option to APG = V &I and Vref(V) = 5, the MAIN PAGE of MASTER will show as Figure
3-46.
As to the voltage setting, the inputted analog voltage 0~5 V maps to the actual output 0~400
V; and for current the inputted analog voltage 0~5 V maps to the actual output 0~60 A as
Figure 3-47(a) shows. Set the APG option to APG = V &I and Vref(V) = 10 means the
inputted analog voltage 0~10 V maps to the actual output 0~400 V for voltage also maps to
the actual output 0~60 A for current as Figure 3-47(b) shows.
Σ
V _ S E T = 4 0 0 . 0 0 V
I _ S E T = 1 5 . 0 0 A
M
S T
O F F
0 0 . 0 0 V
0 0 . 0 0 A
0 . 0 W
A P G - V & I
Figure 3-46