![Chroma 62000P Series Скачать руководство пользователя страница 105](http://html1.mh-extra.com/html/chroma/62000p-series/62000p-series_operating-and-programming-manual_2604534105.webp)
Programmable DC Power Supply 62000P Series Operating & Programming Manual
(3) When the current reaches the set 30V, it remains 10 sec.
(4) Skip to SEQ#3.
(3) SEQ#3:
(1) Since SEQ TYPE = AUTO and TIME=0 are set for SEQ#3, it indicates
SEQ#3 is not executing and the Program is ended.
(4) As RUN COUNT=2 is set, steps (1), (2) and (3) are executed again.
(5) End.
Figure 4-12 shows the output waveform:
COUNT 1
COUNT 2
SEQ#1
SEQ#2
SEQ#1
SEQ#2
START
END
10V
30V
10V
30V
5(Sec)
5(Sec)
10(Sec)
10(Sec)
Figure 4-12
(2) Setting Sequence Type to MANUAL
When SEQ TYPE = MANUAL is set, the Sequence page shown as Figure 4-13 indicates
the sequence will run automatically and stop at the setting of VOLTAGE or CURRENT
without skipping to next sequence until any key on the front panel is presses. It will not
ask users to enter the time the sequence will remain when set to MANUAL.
[ S E Q U E N C E ]
S E Q N O . = 1
S E Q .
T YP E = MA N U A L
V O L T A
G
E
= 0
V S . R .
= 1 . 0 ( V / m S )
C U R R E N T = 0
I
S . R .
= 1 . 0 ( A / m S )
T T L O
U T
= 0
< B I N A R Y = 0 0 0 0 0 0 0 0 >
▲
▼
Figure 4-13
4-12