INST.No.INE-475-P1CE Software Version 1.00
-17-
Setting data name
Display Setting range
of data
Default
value
Remarks
Sample temperature
correction ON/OFF
St ON,
OFF
OFF
The following settings can be executed at "ON".
・
Correction input scaling L
tL
-999 to 999
0
(13)
・
Correction input scaling H
tH
-999 to 999
100
(14)
Surface water ratio
computation ON/OFF
Sd ON,
OFF
OFF
Engineering mode
ENG
The following settings can be executed by pressing [ENT] key.
Processing mode 1:
Weight
α
1
1A
0 to 1.000
0.500
0.5: 3-wavelength processing
0 or 1: 2-wavelength processing
Processing mode 1:
Calibration constant k1
1K
0 to ±9.9999
1.0000
Processing mode 2:
Weight
α
2
2A
0 to 1.000
0.500
0.5: 3-wavelength processing
0 or 1: 2-wavelength processing
Processing mode 2:
Calibration constant k2
2K
0 to ±9.9999
1.0000
Processing mode3:
Weight
α
3
3A
0 to 1.000
0.500
0.5: 3-wavelength processing
0 or 1: 2-wavelength processing
(15)
Processing mode 3:
Calibration constant k3
3K
0 to ±9.9999
1.0000
6. Operation