Rev. 01 of 08/08/2013_CP Page 5 of 15
ASSEMBING OF IR EMITTERS
WARNING
DURING THIS OPERATION THE DEVICE MUST BE DISCONNECTED FROM NETWORK VOLTAGE.
1. Unscrew the two lateral screws of the protection grid;
2. Extract the grid from the left side;
3. Unpack the three IR emitters and screw them on their lamp-sockets;
4. Apply the two black protection on the two ends of the grid and insert the grid into the device;
5. Apply the protection plaque of the grid with the two screws;
At the end of the assembling check what follows :
1. Check tightening of the assembled parts, check if the lamps are well fixed to the fork and that the two
knobs and upper handle are tighted;
2. Check operation of the green warning light: insert the power supply cord plug into a 10A 230V 50Hz
socket and check the lighting of the 230V LINE green warning light;
3. Lighting test:
3.1 – Place on “I” position the three “on/off” switches and load both timers on 15’;
3.2 – Check the lighting of all three IR lamps and the three green lights of the three switches;
3.3 – Check the switching off of the lamps after 15’ when at least one timer is zeroed;
WARNING
DON’T USE THE DEVICE IF ONE OR MORE EMITTERS ARE BROKEN OR MISSED AND/OR THE
PROTECTIVE GRID HAS BEEN REMOVED AND/OR TIMERS ARE BROKEN
Operators that carry out these tests, use the device or are in the environment during lighting must wear
proper goggles.
REQUIREMENTS FOR THE USER
This device can be used only by persons who know operation and hazards associated with the use
and have the following requirements:
Identification Professional user : Doctor or Nurse
a) Age min. 18 years / max 80 years;
b) Education minimum 8 years of schooling but no pre-requisite of specific knowledge;
d) Language Italian or English.