Rev. 01 of 08/08/2013_CP Page 11 of 15
USE BY HANDICAP HOLDERS, AGED AND CHILDREN
It’s necessary the assistance by an operator who knows the functionning and the dangers of the use of this
device.
The device mustn’t be utilized by blind operators.
WARNINGS
KEEP OUT THIS DEVICE FROM CHILDRENS’ REACH.
The professional operators must inform the patients about these directions for use in particular with
regard to:
- Use of eye protection;
- Maintaining the position from the device;
- Contraindications and side effects
MAINTENANCE
IR EMITTERS CLEANING : Frequency of operation
Every three months or before, if you see an opacization of the IR bulbs for dust deposits.
Dust deposits reduce a lot the IR irradiation intensity.
Procedure
Be sure to have disconnected the cable plug from the current socket and that the device is cold.
Remove the clising bracket and extract the protective grid. With a damp cloth clean IR bulbs.
Place the grid again.
ATTENTION
DON'T USE SOLVENTS IN CONTACT WITH PAINTED PARTS AND WHILE WHEELS CLEANING
PAINTED PARTS AND WHILE WHEELS CLEANING: Frequency of operation
Every three months.
Procedure : clean the painted parts with a damp cloth or a cloth imbuted with alcohol. Don’t use solvents.
Remove from the wheels eventual materials (dirtiness, hairs, etc..) embedded in the pins.