Rev. 01 of 08/08/2013_CP Page 13 of 15
IR BULBS REPLACEMENT
ATTENTION
Even if there isn't a great decrease of the radiation intensity of the IR emitters, it's recommended
to replace them after about 4.000 hours of operation.
PROCEDURE
Be sure to have disconnected the plug from the current socket and that the device is cold.
- Remove the protective grid;
- Unscrew the infrared bulb and replace it with a new one;
- Place the grid again.
FUSES REPLACEMENT
For the replacement or the checking of a fuse, open the flap door at the top of the socket block rear the
device; extract the fuses holder inserted into the two boxes marked by the two superior arrows.
ATTENTION
EVERY OTHER REPAIRATION/MAINTENANCE OPERATION MUST BE CARRIED OUT BY AUTHORISED
SERVICE CENTERS FOLLOWING THE PROCEDURES (electric diagrams, bill of materials, descriptions,
test instructions) GIVEN BY THE MANUFACTURER.
Improper modifications of the device and/or intervention by unauthorized persons may result in a risk of
electrical safety for the user.
CAUTION
TO REDUCE THE RISK OF ELECTRIC SHOCK DO NOT OPEN THE DEVICE
INSIDE IT THERE ARE NO PARTS REPARABLE OR REPLACABLE BY THE USER
CONTACT THE MANUFACTURER FOR THE NAME OF THE NEAREST AUTHORIZED SERVICE CENTER
In case of accidental electrical shock, follow the instructions given on the Safety card JE04 “ELECTRIC
SHOCK”