Rev. 01 of 08/08/2013_CP Page 7 of 15
WARNINGS
PLACE THE DEVICE AT LEAST AT ONE METER OF DISTANCE FROM THE WALLS OF THE ROOM.
ALL PERSONS WHO ARE IN THE ROOM DURING LIGHTING MUST WEAR PROTECTIONS AGAINST IR
RADIATIONS.
THE TEMPERATURE IN THE ENVIRONMENT MUSTN’T EXCEED 30°C.
Insert the feeder cable socket into the back plug of the lamp.
Insert the feeder cable plug into a 10A 230V 50Hz socket.
ATTENTION
To avoid electric shock risk, this device must be connected only to supply mains with protection
earth.
The automatic lighting of the green warning light, means that the device is connected to the
feeding network. If it doesn’t light, follow what is described in the section “Breakdowns and
repairs” of this manual.
Set the lamp regulating the wished height and inclination.
Before changing the inclination of the lamps release the knobs at the two ends of the lamp and keep the
lamps using the upper handle.
To avoid the breakage of the bulb or the filament, don't subject the lamp to any knock.
For safety reasons the device is equipped with two series timers. The zeroing of one of the two timers
determines the switching off of the lamp.
Turn carefully in clock-wise sense both the timer knobs and place the arrows in correspondence of the
number of minutes wished for the application. The numbers written around the timers mean “minutes”;
therefore the maximum working time of the device is 15 minutes.
If it's necessary to zero the timers or reduce the programmed time, turn in anti-clock-wise sense the timers
with the maximum care. Rough action on the timers knobs can damage seriously the timers.
ATTENTION
Since the timers have a precision of ± 10% and for safety reasons it’s always necessary that the
time of the application is controlled with a watch apart by a person who attend during the
patient’s treatment.
Light the lamps by pressing the switches “ON/OFF” in correspondence to the “I” (ON) position.
To turn off one or two lamps press the correspondent switch on to "O" (OFF) position.
When the programmed time is over, the device will automatically turn off.
ATTENTION
After the turning off of the emitters, wait at least 10 minutes for the cooling of the device before
a new lighting.
NEVER BLOCK THE FUNCTIONING OF THE TIMERS.
In case of current interruption, when the current is restored the lamps will automatically light if at
least one timer isn’t zeroed.