PERIODIC CHECK OF THE TIGHTINESS OF THE SCHEME ASSEMBLED PARTS: Frequency of operation 
 
Every 3 month or before, if there is anomalous/excessive movement of the connected parts. 
 
Procedure 
 
Check the tightness of the side knobs, the rear fork and the point of connection between base and vertical 
tubes. 
 
 

BREAKDOWNS AND REPAIRS 
 

  ATTENTION 

NO MODIFY OF THIS DEVICE IS ALLOWED; 
BREAKDOWNS AND REPAIRS IMPLEMENTABLE ALSO BY NO-SPECIALIZED PERSONS.

  

 

For safety reasons for every maintenance operation it’s necessary to disconnect the cable plug to the 
network socket. 
 
1° CASE: After inserting the cable plug into the network socket, the green circle warning light doesn’t turn 
on. 
Repair: turn in clock-wise sense both timers place on ”I” position the green switches and then check if the 
IR emitters ignite. 
In case of ignition, call a specialized repairator for the green warning light replacement. 
In case of no ignition, check the wall socket; if it’s functioning check the efficiency of the two fuses placed 
into the socket block back side of the lamp. If they are interrupted, replace them with new fuses 
proceeding as following described. If the fuses are correctly operating, call an authorized service center. 
 
2° CASE: Inserting the cable plug into the network socket, the green warning light turns on, but, after 
having turned both timers and pressed the ON/OFF switches, there isn’t ignition: 
 

-

 

Of all the three IR emitters: call a an authorized service center or Chinesport for the 

reparation. 

-

 

One or two IR emitters: check if the correspondent switches are on the "ON" position and 

their green light is ignited. If so, before replacing the bulbs, verify that they are correctly screwed in 
their lamp holder (for the control and replacement of the bulb proceed as following described). 

 

If the replaced bulbs still doesn’t ignite, it will be necessary to call an authorized service center. 
 
 
 
 

Содержание EL0157

Страница 1: ...Rev 01 of 08 08 2013_CP Page 1 of 15 Item code EL0157 INFRA LAMP 750 Infrared therapy floor lamp User and Maintenance Manual ...

Страница 2: ...ENTS FOR THE USER 05 6 USE OF EYE PROTECTION 06 7 IGNITION 06 8 NOTICES FOR TRANSPORT AND STORAGE 08 9 MEAN OF THE LABELLING SYMBOLS 09 10 SERIAL NUMBER CODIFICATION 09 11 DIRECTIONS FOR USE AND NOTICES 09 12 MAINTENANCE 11 13 BREAKDOWNS AND REPAIR 12 14 USEFUL LIFE OF THE DEVICE 14 15 ENCLOSED ACCESSORIES 14 16 SPARE PARTS 14 Follow us on ...

Страница 3: ...ation of creams by the skin Heating in cold rooms it s possible to obtain an immediate heating for the irradiate person WARNING BEFORE USING THIS MEDICAL DEVICE IT S NECESSARY TO READ THIS MANUAL THAT MUST REMAIN AVAILABLE NEAR THE DEVICE FOR CONSULTATION BY THE OPERATOR EXPOSURE TO INFRARED RAYS HAS CONTROINDICATIONS AND SIDE EFFECTS FOR USE FOLLOW CAREFULLY THE MEDICAL PRESCRIPTIONS TECHNICAL DA...

Страница 4: ...othed washer A and smooth wather B to the two bolts through the holes of tube and base Tight strongly with a spanner Insert the side with stop of the chromated tube into the top end of the painted tube until its stop so that the two holes are lined up String the plated washer into the hexagonal head 8MA bolt Group B insert the bolt with the washer into the two lined up holes and screw it strong wi...

Страница 5: ...z socket and check the lighting of the 230V LINE green warning light 3 Lighting test 3 1 Place on I position the three on off switches and load both timers on 15 3 2 Check the lighting of all three IR lamps and the three green lights of the three switches 3 3 Check the switching off of the lamps after 15 when at least one timer is zeroed WARNING DON T USE THE DEVICE IF ONE OR MORE EMITTERS ARE BRO...

Страница 6: ...N Replace these protections in case of breakage Clean before every use with neutral a diluited sterilizing product The materials used in these eye protections may cause allergic reactions to sensible persons IT S NECESSARY TO USE EYE PROTECTIONS ACCORDING THE ABOVE MENTIONNED STANDARDS IGNITION Before lighting read these important warnings WARNINGS THIS MEDICAL DEVICE MUST BE USED ONLY BY MEDICAL ...

Страница 7: ...e of the two timers determines the switching off of the lamp Turn carefully in clock wise sense both the timer knobs and place the arrows in correspondence of the number of minutes wished for the application The numbers written around the timers mean minutes therefore the maximum working time of the device is 15 minutes If it s necessary to zero the timers or reduce the programmed time turn in ant...

Страница 8: ... Store in a clean and dry place ATTENTION During the therapeutic session it s necessary the control by the operator to avoid the risk that the patient moves from his position removes his eye protection or act on the control panel NOTICES FOR TRANSPORT AND STORAGE During the transport it s necessary to avoid shocks on the packaging containing the IR lamps In case of damaged packaging it s necessary...

Страница 9: ... the following mean The first two figures represent the year of production The following three figures represent the serial number DIRECTION FOR USE AND NOTICES Infrared therapy is a physiotherapeutic technique that uses for therapeutic purposes the effects produced by the infrared rays in the tissues The main biological effect of infrared rays is the thermal effect The infrared rays produce heat ...

Страница 10: ...anger of burning the same After treatment do not go out to drafts or sudden changes of temperature Dangerousness connected to the utilization of these burners is mainly due to HEAT ABSORPTION BY THE EYE Extremely high irradiation power in every wave lenght causes damages to the human eye due to the absorption of the developed heat with time years or some decades it can make muddy the optical lens ...

Страница 11: ...Frequency of operation Every three months or before if you see an opacization of the IR bulbs for dust deposits Dust deposits reduce a lot the IR irradiation intensity Procedure Be sure to have disconnected the cable plug from the current socket and that the device is cold Remove the clising bracket and extract the protective grid With a damp cloth clean IR bulbs Place the grid again ATTENTION DON...

Страница 12: ... call a specialized repairator for the green warning light replacement In case of no ignition check the wall socket if it s functioning check the efficiency of the two fuses placed into the socket block back side of the lamp If they are interrupted replace them with new fuses proceeding as following described If the fuses are correctly operating call an authorized service center 2 CASE Inserting t...

Страница 13: ...act the fuses holder inserted into the two boxes marked by the two superior arrows ATTENTION EVERY OTHER REPAIRATION MAINTENANCE OPERATION MUST BE CARRIED OUT BY AUTHORISED SERVICE CENTERS FOLLOWING THE PROCEDURES electric diagrams bill of materials descriptions test instructions GIVEN BY THE MANUFACTURER Improper modifications of the device and or intervention by unauthorized persons may result i...

Страница 14: ...commended to keep the counting of the operating hours and periodically replace the emitters ACCESSORIES AND SPARE PARTS REPLACEMENT ATTENTION Accessories and spare parts must correspond to what indicated in this manual and used by the manufacturer The replacement with different component could compromise the level of safety of the device or cause unintended effects on the patient and or operator T...

Страница 15: ...e Before contacting the Service Center please identify the SERIAL NUMBER in the product label attached to the equipment see facsimile above or the transport document invoice number This will speed things up and improve service quality A piece of equipment that requires repairs at Chinesport s facilities must have an AUTHORIZATION NUMBER on the shipment packaging This number must be requested from ...

Отзывы: