–11 –
Measuring:
The display now shows the measured tension values.
To remove the material to be measured:
1. Press the lever and remove the material to be measured.
2. Slowly release pressure on the lever until the guide rollers return to their original
position.
5.3 Damping
This feature is used for tensions that vary strongly
In the Damping menu, you can specify separate damping factors for each material
characteristic.
1. Back in the display mode, press the
button to activate or deactivate the damping
function. This is recommended if the displayed values vary strongly.
2. Press the button to increase the damping value or the button to decrease it.
Changes made to the damping factor using the arrow keys are not saved
permanently in the material characteristic. If you disable the damping feature
or select a different material characteristic, the damping factor will be reset to
the value originally saved for the material characteristic.
The factory setting for the damping factor is 5. The average shown on the display is
calculated as follows:
The factory setting for the damping factor is 5. The average shown on the display
is calculated as follows:
5 old measured 4 new measured values
9
Damping can be changed in 9 steps from 01 = low damping
1 old measured 8 new measured values
9
to 9 = high damping
8 old measured 1 new measured values
9
Содержание ETB