– 9 –
4.1 Material menu
In the material menu you can make the settings for the selected material characteristics
and perform the calibration. To perform the calibration, the weights for the selected
calibration points must be available.
Material Setup
Submenu
Settings Menu
Description
No. of the charac-
teristic material-
curve 1 - 4
–
[character]
[numbers]
[special character]
to enter a name for the selected material char-
acteristic. The name can also been adjusted by
using the provided software
Damping
–
[1] – [9]; [5]
Section 5.3
Alarms
–
[ON], [OFF]
To activate or deactivate the alarm
High Limit
–
[0000], [9999]
If the set limit value is exceeded, the display
reads MAX-ALARM.
Low Limit
–
[0000], [9999]
If the value falls below the set limitvalue, the
display reads MIN-ALARM.
Calibration Chapter
3.6.2
Start
To perform a calibration, follow the instructions
shown on the display.
Cal Points
between [5%] and [90 %] of
the tension range
Set three calibration points for which a
calibration should be performed.
Weight
[cN], [g]
Set the unit of the calibration weights used.
Factory Settings are illustrated in bold
4.2 Password
By setting a password, you can block the access to some menu parts for unauthorized
users. There are 3 protection levels:
0 All menus are accessible (Factory setting)
1 The system setting menu is locked
2 The material, system setting, memory setting and cal. adjustment menus are locked
3 All menus are locked.
To apply the password protection
1. Select the menu password.
2. Insert a password and choose a protection level.
3. Exit the main menu, to apply the password protection.
NOTE:
The factory setted password is 0000. If you forgot the password, please contact
the factory to request the master password.
4.3 Factory reset
A factory reset resets the tension meter to its original manufacturer settings.
This procedure will delete all settings, including any customer-defined material
characteristics (calibrations); the factory calibration, however, will be kept.
WARNING:
Customer calibrations will be deleted.