– 10 –
5.0 OPERATION PROCEDURES
Requirements:
1. Switch the tension meter on (section 3.2)
2. Define the required tension meter settings (section 4.0)
3. Select the desired material characteristic (section 4.1)
4. Bring the tension meter into the desired measuring position and carry out a zero
adjustment as described in section 5.1, if required.
5.1 Zero adjustment of the measurement position
Each time the measurement position is changed, the tension meter will
automatically perform a zero adjustment.
If the tension meter does not display zero in its measuring position, perform a
manual zero adjustment procedure. For this purpose, no material to be measured
must have been inserted yet!
To carry out zero adjustment:
1. Switch on the tension meter
2. Bring the tension meter into the desired measuring position and make sure not to
move it.
3. Press the
button.
The Display momentarily shows
and then switches to
. The
tension meter is now adjusted to its new measuring position and ready to take
measurements.
5.2 Inserting and removing material to be measured
Inserting the material to be measured:
1. Press the lever to tilt the outer guide
rollers sidewards.
2. Thread the material to be measured
through the measuring and guide rollers
(filament guide).
3. Slowly release pressure on the lever
until the guide rollers return to their
original position.
NOTE:
It is important to assure that the
material to be measured runs smoothly
between the measuring and guide rollers.