Step 1. Set the calibration points and weights, e.g. in Newton using the
calibration points 10%, 40%, and70 % of full scale
Start: Select the Start menu item.
Alternative calibration points: 5%, 45%, 90% 10%, 50%,90 %
5 %, 50%, 90% 10 %, 40%, 70 %
10%, 45%, 90%
We recommend to use the 10%, 50%, 90% setting for an initial calibration.
If the measured values are not accurate enough when you check the
calibration, repeat the calibration using other calibration points. We
recommend that you align the middle of the tension range to be measured
with the middle calibration point.
Step 2. Perform a zero adjustment with the tension meter in its measuring
position
Step 3: Calibrate calibration point 1 Insert material to be measured
Step 4: Calibrate calibration point 2
Step 5: Calibrate calibration point 3
NOTE:
Once you have completed the calibration, make sure to verify it as described in
sectio 6.3. In case of a large deviation, repeat the calibration or select different calibration
points.