51
15000V range
V
0
2000
4000
6000
8000
10000
12000
14000
16000
1,E-02
1,E-01
1,E+00
1,E+01
1,E+02
1,E+03
1,E+04
1,E+05
1,E+06
1,E+07
1,E+08
10 k 100 k 1 M 10 M 100 M 1 G 10 G 100 G 1 T 10 T 100 T
W
10 k 100 k 1 M 10 M 100 M 1 G 10 G 100 G 1 T 10 T 100 T
8.2.4. dAr, PI, ANd dd
Calculation of the dAr and PI terms
Specified range
0.02 … 50.00
Resolution
0.01
Intrinsic uncertainty
± (5% + 1 pt)
Calculation of the dd term
Specified range
0.02 … 50.00
Resolution
0.01
Intrinsic uncertainty
±( 10% + 1 pt)
8.2.5. CAPACItANCE
Capacitance measurement
This measurement is made following the discharging of the element tested, after each measurement
Specified
measurement range
0.005 … 9.999µF
10.00 … 49.99µF
Resolution
1 nF
10 nF
Intrinsic uncertainty *
± (10% + 1 pt)
± 10%
*: this uncertainty is specified only for a test voltage
≥
500V.