36
DISPLAY
Press the DISPLAY key to see
the following screen.
Test Timing
Timed Run (m:s)
DAR (s/s)
PI (m/m)
2:00
30/60
1.0/10
timed run
: to set the measurement duration (in minutes:seconds) for measurements with programmed duration.
The adjustment range is from 00:10 to 99:59, in 1-second steps.
dAr
: to set the time at which the measurements must be recorded to calculate the DAR. This can be used in special applications.
The first time can be set from 10 to 90 seconds in 5-second steps.
The second time can be set from 15 to 180 seconds in 5-second steps.
PI
: to set the time at which the measurements must be recorded to calculate the PI. This can be used in special applications.
The first time can be set from 0.5 to 30 minutes in 0.5- then 1-minute steps.
The second time can be set from 0.5 to 90 minutes in 0.5-, 1-, and 5-minute steps.
5.2. GENErAL PArAMEtErS
buzzer
: to set the audible level of beeps: 1, 2, 3, or Off (no sound).
Power down
: automatic switching off of the instrument: On (switching off at the end of 5 minutes), Off (no switching off).
baud rate
: to set the data rate to 9600, 19200, 38400 or 57600 bauds.
date
: to set the date in yyyy-mm-dd format.
time
: to set the time in hh:mm format.
temperature unit
: to choose the unit of temperature: Celsius or Fahrenheit.
Instrument Number
: indicates the number of the instrument. This line is informative and cannot be modified.
firmware
: indicates the software version of the instrument. This line is informative and cannot be modified.
5.3. MEASUrEMENt PArAMEtErS
DISPLAY
Press the DISPLAY key to see
the following screen.
Test Parameters
Test Type
Maximum Output Current
Maximum Output Voltage
Adjustable Voltage 1
Adjustable Voltage 2
Adjustable Voltage 3
Burn-in
5.0 mA
15000 V
50 V
500 V
2500 V