3.6.2. INtErPrEtAtIoN of tHE rESULt
Value of dd
Quality insulation
7 < DD
Very poor
4 < DD < 7
poor
2 < DD < 4
Borderline
DD < 2
Good
3.7. CAPACItANCE MEASUrEMENt
The capacitance measurement is made automatically during the insulation measurement, and is displayed after the measurement
has been stopped and the device tested has been discharged.
3.8. MEASUrEMENt of tHE rESIdUAL CUrrENt
The measurement of the residual current flowing in the device tested is made automatically as soon as the connection to the
device tested is made, then during and after the insulation measurement.