93
APPENDIX
EN
INTELECT
®
RPW 2 USER MANUAL
ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES (CONTINUED)
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Intelect
®
RPW 2 is intended for use in the electromagnetic environment specified below. It is the responsibility of the customer or user to ensure
that the Intelect RPW 2 device is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Conducted RF
to IEC 61000-4-6
6V
0,15MHz-80 MHz
80% AM at 1kHz
6V
0,15MHz-80 MHz
80% AM at 1kHz
Radiated RF
to IEC 61000-4-3
10 V/m
80MHz-2,7GHz
80% AM at 1kHz
10 V/m
80MHz-2,7GHz
80% AM at 1kHz
Proximity fields from RF
wireless communications
equipment
to IEC 61000-4-3
27 V/m
380-390MHz
PM 18Hz
28 V/m
430-470MHz
FM +/- 5kHz
1 kHz sine
9 V/m
704-787 MHz
PM 217Hz
28 V/m
800-960 MHz
PM 18 Hz
28 V/m
1,7-1,99 GHz
PM 217 Hz
28V/m
2,4GHz-2,57GHz
at PM 217Hz
9V/m
5,1GHz-5,8GHz
at PM 217Hz
27 V/m
380-390MHz
PM 18Hz
28 V/m
430-470MHz
FM +/- 5kHz
1 kHz sine
9 V/m
704-787 MHz
PM 217Hz
28 V/m
800-960 MHz
PM 18 Hz
28 V/m
1,7-1,99 GHz
PM 217 Hz
28V/m
2,4GHz-2,57GHz
at PM 217Hz
9V/m
5,1GHz-5,8GHz
at PM 217Hz
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Intelect RPW 2 exceeds the applicable RF compliance level above, the Intelect RPW 2 device should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the Intelect RPW 2 device.
Содержание Intelect RPW 2
Страница 32: ...32 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Select transmitter ...
Страница 43: ...43 USER INTERFACE EN INTELECT RPW 2 USER MANUAL 7 START TREATMENT Press the START button Treatment Running ...
Страница 48: ...48 USER INTERFACE EN INTELECT RPW 2 USER MANUAL 2 DELETE Treatment Data Delete all IDs Delete one ID ...
Страница 49: ...49 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Confirm or cancel Delete all treatment sessions ...
Страница 50: ...50 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one session ...
Страница 55: ...55 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one protocol ...
Страница 61: ...61 USER INTERFACE EN INTELECT RPW 2 USER MANUAL 4 Full screen image Close full screen mode ...