76
DEVICE MAINTENANCE
EN
INTELECT
®
RPW 2 USER MANUAL
CLEANING THE INTELECT
®
RPW 2
With the system disconnected from the power source, clean
the system with a clean, lint-free cloth moistened with water
and mild antibacterial soap. Do not use solvents. If a more
sterile cleaning is needed, use a cloth moistened with an
antimicrobial cleaner. Cleaning should be performed daily.
Do not submerse the system in liquids. Should the device
accidentally become submersed, contact the dealer or DJO
Service Department immediately. Do not attempt to use a
system that has been wet inside until inspected and tested by
a Trained Technician.
Cleaning the LCD Screen
Clean the LCD with a clean, dry cloth, in the same way as
cleaning the computer monitor screen. Do not use abrasive
materials or chemicals or liquids.
COPY OF MANUAL
To obtain a copy of the Intelect RPW 2 User Manual, contact
your local representative or DJO Global
®
Customer Care
INSTRUCTION FOR SOFTWARE UPGRADE
1. Download firmware upgrade from the Chattanooga
®
website www.chattanoogarehab.com onto the USB stick.
2. Switch OFF the device
3. Insert USB key
4. Switch ON the device
5. Device will automatically detect firmware update availability
6. Confirm firmware update
DEVICE MAINTENANCE
No internal maintenance or routine calibration is required for
the device itself.
For the Falcon
®
and V-Actor HF
®
handpieces along with the
transmitters, please refer to the correct section in this user
manual.
Power cable replacement: Unplug the default cable, then plug
the new cable.
*Falcon and V-Actor are registered trade marks of Storz medical AG
DEVICE MAINTENANCE
WATER RESERVOIR
The RPW 2 uses a condenser between the compressor
and the handpiece to extract air humidity to avoid water
accumulation in the hand piece. The extracted water is
collected in the water reservoir located on the back panel
just below the power cord port.
The water level should be checked every day before use. The
water reservoir should be emptied on a regular basis.
To empty the reservoir follow the instructions.
1. Press the release clip and gently lift-up the water reservoir
from the Intelect
®
RPW 2, sliding it up.
2. Remove the lid of the water reservoir carefully and
dispose of the waste water.
To reattach the reservoir
1. Replace the reservoir lid and insert the water tube into the
opening on the top
2. Place the reservoir against the Intelect RPW 2 above the
connection clip
3. Slide the reservoir down. Press down gently and there will
be an audible click. The release clip will still be visible after
the water reservoir is installed.
Содержание Intelect RPW 2
Страница 32: ...32 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Select transmitter ...
Страница 43: ...43 USER INTERFACE EN INTELECT RPW 2 USER MANUAL 7 START TREATMENT Press the START button Treatment Running ...
Страница 48: ...48 USER INTERFACE EN INTELECT RPW 2 USER MANUAL 2 DELETE Treatment Data Delete all IDs Delete one ID ...
Страница 49: ...49 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Confirm or cancel Delete all treatment sessions ...
Страница 50: ...50 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one session ...
Страница 55: ...55 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one protocol ...
Страница 61: ...61 USER INTERFACE EN INTELECT RPW 2 USER MANUAL 4 Full screen image Close full screen mode ...