SERVICE AND WARRANTY
EN
INTELECT
®
RPW 2 USER MANUAL
SERVICE LIFE OF THE FALCON
®
HANDPIECE
The Falcon
*
handpiece should be overhauled after around
every 1 million shocks. Provided this interval is observed, the
average expected service life is approx.
• 5 million shocks for the handpiece
• 1 million shocks for the shock transmitters
• 5 million shocks for the clutch.
Exceeding the service life can be expected to result in a
failure of the devices. No warranty claims shall be accepted
beyond the information given.
SERVICE LIFE OF THE V-ACTOR HF
®
HANDPIECE
The average expected service life of the handpiece is approx.
5 million pulses. Exceeding the service life can be expected
to result in a failure of the devices.
No warranty claims shall be accepted beyond the
information given in the warranty section of this manual.
*Falcon and V-Actor are registered trade marks of Storz medical AG