![Chattanooga Group Intelect Legend XT Скачать руководство пользователя страница 75](http://html.mh-extra.com/html/chattanooga-group/intelect-legend-xt/intelect-legend-xt_service-manual_2592340075.webp)
70
Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
6 REMOVAL/REPLACEMENT
6.6 THERAPY SYSTEM KEYMAT ASSEMBLY
A. Tools and Equipment Required
#1 Phillips Screwdriver
Needle Nose Pliers
B. Keymat Assembly Removal
Separate Top from Bottom.
Refer to 6.3,
part B.
Remove Fan.
Refer to 6.4, part B.
Remove Control Board Assembly.
Refer to
6.5, part B
.
Remove eight Keymat PC Board Retaining
Screws.
See Figure 6.28
.
Starting at the top of the Keymat PC Board,
lift board with one hand, push in and release
the plastic clips holding the Keymat PC Board
in position.
See Figure 6.29
.
Remove Keymats from the System Top.
See
Figure 6.30.
C. Replacing Keymat Assembly
Replace with new Keymat Assembly Kit by
reversing the
steps 1- 6
above.
NOTE:
Do not over tighten the screws. Over
tightening will damage the threads of the
brass inserts.
Re-assemble Therapy System referring to
the appropriate sections of this manual for
proper instructions.
1.
2.
1.
2.
3.
4.
5.
6.
1.
2.
FIGURE 6.28
FIGURE 6.29
REMOVE
RETAINING
SCREWS
LIFT PC BOARD AND
RELEASE PLASTIC CLIPS
FIGURE 6.30
REMOVE ALL THREE
KEYMATS
DISCONNECT THE SYSTEM FROM THE POWER
SOURCE (OUTLET OR REMOVE BATTERY MODULE IF
INSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE,
INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM.
MAKE CERTAIN WHEN INSTALLING CONTROL BOARD
ASSEMBLY TO KEYMAT PC BOARD PINS FROM THE
CONTROL BOARD ALIGN WITH THE CONNECTOR ON
THE KEYMAT PC BOARD.