Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
5 TROUBLESHOOTING
5.10 RUSSIAN MODE TEST
Set up System per
5.6 parts A and B
prior to
performing test.
A. Russian Mode Test Procedures
Set Scope; Time- 5 mS, Channel- 50 V, and
Trigger- DC.
NOTE:
A test of the Optional Patient Interrrupt Switch
is provided within the Russian Mode Test. If
you do not have the Optional Patient Interrupt
Switch, skip
steps 2, and 11
.
Install Patient Interrupt Switch.
See
Figure 5.8.
Press Electrotherapy Button.
Press Russian Button. Press Edit Button.
Press Channel Mode until Co-Contract is
displayed.
Press the Cycle Time Button until
Continuous is displayed.
Rotate Therapy Intensity Control clockwise
until 100 is displayed.
Press “Start”.
Compare waveform on scope to
Figure 5.9.
Verify that both Channels reach 100.
Press Patient Interrupt Switch. Verify
treatment stops and “Patient Switch for
Ch 1 and 2 was pressed. Press any button
to continue...” message appears.
See
Figure 5.9A.
Press any button.
Press Stop Button. Press Home Button and
move scope probes to Channel 3 and 4
on 4 Channel Systems. Repeat
steps 2
through 11
.
B. Russian Mode Test Results
Waveform is the same between scope
and Figure 5.9, amplitude reached 100
and patient switch message displayed
when switch pressed.
See Figure 5.9A
.
Unit passed test.
No waveform or considerably different
waveform.
Unit failed test. Replace appropriate Stim
Board.
Amplitude failed to reach 100 on both
Channels.
Unit failed test. Replace appropriate Stim
Board.
Patient Switch message did not display
when patient switch pressed.
(Optional Patient Switch present)
Unit failed test.
a)
Try a known good Patient Switch and
repeat
test.
b)
Replace appropriate Stim Board.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
1.
2.
3.
4.
FIGURE 5.8
FIGURE 5.9
FIGURE 5.9A