Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
5 TROUBLESHOOTING
ERROR MESSAGES REQUIRING TECHNICAL ASSISTANCE 200299
Code
Number
Type
Message
Probable Cause
Possible Remedies
200
Error
Error reading the system Real Time Clock (RTC)
Replace Control Board
201
Error
Internal List Box Memory Error.
Reinstall software. If problem persists, replace Control Board.
202
Error
Program Control Software Allocation Memory Error
Reinstall software. If problem persists, replace Control Board.
203
Error
Error erasing Patient Data Card
Insert known good Patient Data Card. If problem persists, replace Control Board.
204
Error
Error writing to sEMG Data Card.
Insert known good sEMG Data Card. If problem persists, replace Control Board.
205
Error
MMC Card Formatting Error.
(Using Tech Service Utilities)
Insert known good MMC Card. If problem persists, replace Control Board.
206
Error
Error reading MMC.
Insert known good MMC Card. If problem persists, replace Control Board.
207
Error
Error reading protocols on power up of system.
Restore default protocols.
All User Protocols and Saved Sequences will be deleted
permanently from the system.
208
Error
Error reading protocols.
Restore default protocols.
All User Protocols and Saved Sequences will be deleted
permanently from the system.
209
Error
Error writing protocol.
Restore default protocols.
All User Protocols and Saved Sequences will be deleted
permanently from the system.
210
Error
Error Calibrating Ultrasound Applicator.
A.
Connect a known good Ultrasound Applicator.
B.
Replace Ultrasound Board.
C.
Replace Control Board.
211
Error
Error saving Calibration Data to Ultrasound Applicator.
A.
Connect a known good Ultrasound Applicator.
B.
Replace Ultrasound Board.
C.
Replace Control Board.
212
Error
Ultrasound Applicator not calibrated OK Error.
A.
Connect a known good Ultrasound Applicator.
B.
Replace Ultrasound Board.
C.
Replace Control Board.
213
Error
Time out error saving Ultrasound Applicator
Calibration Data to Applicator.
A.
Connect a known good Ultrasound Applicator.
B.
Replace Ultrasound Board.
C.
Replace Control Board.
214
Error
General Laser PC Board Error.
Send Laser Module to factory for Service.
215
Error
Laser Applicator out of calibration due to laser output
being too high.
Send Laser Applicator to factory for Service.
216
Error
Laser Applicator out of calibration due to laser output
being too low.
Send Laser Applicator to factory for Service.
217
Error
Laser Applicator out of Calibration Due to LED output
being too high.
Send Laser Applicator to factory for Service.
218
Error
Laser Applicator out of Calibration Due to LED output
being too low.
Send Laser Applicator to factory for Service.
219
Error
Error while performing a Software upgrade.
A.
Turn Therapy System Off and back On. Reattempt upgrade.
B.
Replace the PC Board that the software is attempting to upgrade.
220
Error
Error while performing a Software upgrade.
A.
Turn Therapy System Off and back On. Reattempt upgrade.
B.
Replace the PC Board that the software is attempting to upgrade.
221
Error
Error while performing a Software upgrade.
A.
Turn Therapy System Off and back On. Reattempt upgrade.
B.
Replace the PC Board that the software is attempting to upgrade.
222
Error
Error while performing a Software upgrade.
A.
Turn Therapy System Off and back On. Reattempt upgrade.
B.
Replace the PC Board that the software is attempting to upgrade.
223
Error
Error while performing a Software upgrade.
A.
Turn Therapy System Off and back On. Reattempt upgrade.
B.
Replace the PC Board that the software is attempting to upgrade.
5.1 THERAPY SYSTEM ERROR MESSAGES (CONTINUED)