![Chattanooga Group Intelect Legend XT Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/chattanooga-group/intelect-legend-xt/intelect-legend-xt_service-manual_2592340028.webp)
23
Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
4 SPECIFICATIONS
*CC= Constant Current
CV= Constant Voltage
FIGURE 4.10
I.
VMS™ Burst- Figure 4.10 (Vectra Genisys)
Output Mode . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . .0-200 mA (CC)
0-200 V (CV)
Channel Mode . . . . Single, Reciprocal, Co-Contract
Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . 20-400 μsec
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Anti-Fatigue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Off or On
Set Intensity . .Individual Channel Intensity Setting
in Reciprocal and Co-Contract modes
Cycle Time . . . . . . . . . . . . . . . . . .Continuous, 5/5, 4/12,
10/10, 10/20, 10/30, 10/50
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 pps
Ramp . . . . . . . . . . . . . 0.5 sec, 1 sec, 2 sec, and 5 sec
Treatment Time . . . . . . . . . . . . . . . . . . 1-60 minutes
Available on Channels . . . . . . 1 & 2, 3 & 4 Option
J. DC (Direct Current)- Figure 4.11 (Vectra
Genisys)
Output Mode . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . 0-4 mA
Polarity Reversal . . . . . . . . . . . . . . . . . . . . . .On or Off
With Polarity Reversal On, Polarity will change
after 50% of treatment time.
Cycle Time . . . . . . . . . . . . Continuous, 5/60, 10/60
Treatment Time . . . . . . . . . . . . . . . . . . 1-10 minutes
Mode Selection (Fixed) . . . . . . . . . . . . . . . . . . . . . . . . . CC*
Available on Channels . . . . . . 1 & 2, 3 & 4 Option
K. Iontophoresis- (Intelect Vet)
Output Mode . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . 0-4 mA
Calculated Dosage . . . . . . . . . 40-80 mA - Minute
Mode Selection (Fixed) . . . . . . . . . . . . . . . . . . . . . . . . . CC*
Available on Channels . . . . . . 1 & 2, 3 & 4 Option
4.2 ELECTROTHERAPY WAVEFORM SPECIFICATIONS (CONTINUED)
FIGURE 4.11