32
Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
5 TROUBLESHOOTING
Code
Number
Type
Message
Probable Cause
Possible Remedies
135
Warning
Control Board Software upgrade warning.
Upgrade Control Board Software to latest version. Contact dealer or Chattanooga Group for latest
software upgrade and instructions.
136
Warning
Stim Board Main Software upgrade warning.
Upgrade Stim Board Software to latest version. Contact dealer or Chattanooga Group for latest
software upgrade and instructions.
137
Warning
Stim Board Main Software upgrade warning.
Upgrade Stim Board Software to latest version. Contact dealer or Chattanooga Group for latest
software upgrade and instructions.
138
Warning
Ultrasound Board Software upgrade warning.
Upgrade Ultrasound Board Software to latest version. Contact dealer or Chattanooga Group for
latest software upgrade and instructions.
139
Warning
Laser Board Software upgrade warning.
Upgrade Laser Board Software to latest version. Contact dealer or Chattanooga Group for latest
software upgrade and instructions.
140
Warning
MMC Software upgrade warning.
Upgrade MMC Software to latest version. Contact dealer or Chattanooga Group for latest software
upgrade and instructions.
141
Warning
Battery Module Software upgrade warning.
Upgrade Battery Software to latest version. Contact dealer or Chattanooga Group for latest
software upgrade and instructions.
142
Warning
A Laser Protocol was selected but no Laser Module is
installed on system.
Install Laser Module to Therapy System. Refer to Laser Module User Manual for installation
Instructions.
143
Warning
A Laser Protocol was selected but no Laser Applicator
connected to system.
A.
Connect proper Laser Applicator to the system.
B.
If Laser Applicator is connected, reset Therapy System by turning main power switch
Off and On.
C.
Connect a known good Laser Applicator.
D.
If problem continues, send Laser Module to factory for Service.
144
Warning
Wrong Laser Applicator connected to system for the
protocol selected.
A.
Connect correct Laser Applicator to the system.
B.
If Applicator is connected, reset Therapy System by turning main power switch Off and On.
C.
Connect a known good Laser Applicator.
D.
If problem continues, send Laser Module to factory for Service.
145
Warning
Patient Data Card button on Home Screen was pressed
with no Patient Data Card installed into system port and
no treatment currently being performed.
Properly insert a Patient Data Card, set up and perform the treatment and, save data to Patient
Data Card.
5.1 THERAPY SYSTEM ERROR MESSAGES (CONTINUED)