Endosafe
®
-
Portable Test System User’s Guide
Copyright 2003
Page D3
PIPTS101-10
RETRIEVING RESULTS OPTIONS
1.
You can download results directly to your PC and retrieve from the
designated location file.
2.
You can use the Seiko
®
DPU-414 or the Epson TM-U220D Printer
(available from Charles River Laboratories) to print the last test result, all
results from a particular date, or up to a maximum of one hundred (100)
stored test results.
Example of Glucan Test Report:
********* ENDOSAFE Test Record *********
V7.12B 12/9/2008
DateTime: ........ 04/21/09 @ 08:59:40AM
Device: ........................... 0633
OperatorID: ......................... DN
Cartridge: .................. PTS Glucan
Temperature: .. Start: 37.0C End: 37.0C
Method: ......................... KX-122
Cartridge Lot#: ................ 9107441
Cartridge Cal Code: ....... 901630102934
Range: ..................... 1000-10.00
Range Time: ................ Sec: 16-701
Onset Times: ........ >701 098 >701 102
Slope: -0.821 ........ Intercept: +3.667
Dilution: ............................ 1
Sample Lot: ................... 21042009
Sample ID: ...................... RD0901
Sample Rxn Time CV: .......... 0.0% Pass
Spike Value: ............... 97.2 pg/mL
Spike Rxn Time CV: ........... 2.8% Pass
Spike Recovery: ............. 99% Pass
Test Suitability: ................. Pass
Sample Value: .............. <10.0 pg/mL
: